Setting Your Saniclave 200.
The sterilization cycle parameters are as follows: 30 minutes, 121C, 15psi. The
sterilization cycle parameters cannot be adjusted.
Load Settings
Load type
Wrapped
Loads
Unwrapped
Loads
Note: For wrapped loads, the user must use wraps or pouches that are
cleared by the FDA for human use. Unwrapped loads must be used
immediately upon sterilization.
Set Time and Date – To set the time and date, press and hold the 'SET'
button. The 'ADVANCE' button may be used to advance from the time,
date and year. The 'START' and 'SET' buttons can be used to
settings up and down. Disconnecting the power between uses may
interrupt the clock's timing. Date and time may need to be reset. If set
correctly, the time and date will be printed on the printout.
Print – After a cycle is complete, when display reads 'CC,' press the SET button
once. Other than the printouts, the Saniclave 200 does not record cycle
data.
See page 6 for more printer information.
R E V O L U T I O N A R Y S C I E N C E
Instructions
Press the WRAPPED button. The display should change from
0000 to 'C1' and the preheat mode will immediately follow.
After the sterilization mode is complete, the display will read
'turn door.' At this time, remove the door and insert the
bottom edge under the retainer clips. Press ADVANCE to
begin the dry mode. When the dry mode ends, the unit will
read 'CC'
This setting employs the dry mode .
Press the UNWRAPPED button. The display should change
from 0000 to 'C2' and the preheat mode should commence.
When the sterilization cycle ends, the unit will read 'CC'
This setting does not employ the dry mode.