Introduction Dear Valued Customer, Thank you for purchasing the Revolutionary Science Saniclave 200. We are confident that the Saniclave 200 will provide years of reliable service. The high quality of construction and durable parts deliver a performance equivalent to much costlier equipment.
Shafer, MN 55074. Intended Use The Revolutionary Science Saniclave 200 is designed to be used in medical and dental clinics, hospitals and other facilities where reusable sterile equipment is used. It is intended to sterilize wrapped and unwrapped heat and moisture stable solid...
Page 7
Saniclave RS-SC-200 Limitations of Use Revolutionary Science is not responsible for misuse of the Saniclave 200 1. NEVER attempt to dismantle or modify the Saniclave 200. 2. NEVER leave operating Saniclave 200 unattended. 3. NEVER block drain (located on the inside of the chamber below heating element).
Saniclave RS-SC-200 Operating the Saniclave 200 Preparation for Sterilization (Default mode) 1. Preclean instruments according to their manufacturer's recommendations. 2. Follow instrument manufacturer’s instructions for correct cleaning and lubricating procedures. WARNING! Failure to properly clean instruments before autoclaving, may result in unsterile equipment. Using unsterile equipment may result in serious injury or death.
Page 9
Saniclave RS-SC-200 Operating Instructions 1. Place Saniclave 200 on a dry, level surface. 2. Remove chamber door by rotating door counter clockwise. 3. Fill chamber with approximately 500mL of distilled water (until the heating element is fully submerged and the metal portion of the temperature sensor is in contact with the water).
Page 10
Saniclave RS-SC-200 17. Dry Mode: After the chamber drains, the autoclave will sound an audible alarm that reads “turn door.” Rotate door counter clockwise until it is in the open position. Tuck the door under the retainer clips below the door opening.
Page 11
Saniclave RS-SC-200 Setting Your Saniclave 200. The sterilization cycle parameters are as follows: 30 minutes, 121C, 15psi. The sterilization cycle parameters cannot be adjusted. Load Settings Load type Instructions Wrapped Press the WRAPPED button. The display should change from Loads 0000 to 'C1' and the preheat mode will immediately follow.
Troubleshooting Error Conditions and Display Readings Note: If a failure occurs, the load must be considered non-sterile. The sterilizer and load should then be investigated. Contact Revolutionary Science if problem persists. 1. Manual Cycle Abort: If 'START' is pressed, display will read 'Err 1' for cycle abort.
Page 13
'drn tank full.' At this point, the exhaust tank must be emptied if the end of the drain tube is not already rest - ing in a sink or bucket. Revolutionary Science recom mends that the drain tube be left in the sink or other reservoir for regular drainage.
_______________________________________________________________ Warranty Your REVOLUTIONARY SCIENCE Saniclave 200 is warranted to be free from defects of material and workmanship under normal use for a period of two years from date of purchase. If your Saniclave 200 proves to be defective within this warranty period, APR&D, Inc.
CE / IEC Safety: RS-SC-200 tested to and found to be compliant with IEC 61010A and associated safety standards. ISO 9001: Revolutionary Science is an ISO Registered Company IQ/OQ/PQ: Revolutionary Science provides a range of compliance statements and IQ/OQ/PQ documents that will enable customers to successfully validate their products in accordance with regulations and standards.
_______________________________________________________________ _______________________________________________________________ _______________________________________________________________ Revolutionary Science recommends the weekly use of G. Bacillus Stearothermophilus biological indicator ampules. Place indicator in center of full load and test weekly. The use of FDA cleared class 5 integrators should be used with each load. Follow the manufacturer's recommendations for class 5 integrator use.
All repairs must be performed by a trained professional. Contact Revolutionary Science for more information. (See the contact section on page 12) R E V O L U T I O N A R Y S C I E N C E...
Page 18
Saniclave RS-SC-200 Saniclave 200 Cycle Documentation Table It is recommended to keep records of each sterilization cycle. This table may be copied. R E V O L U T I O N A R Y S C I E N C E...
Saniclave RS-SC-200 Sterilization Monitoring Guidelines Note Revolutionary Science provides the following information only for reference. The user must contact state and local government for proper regulations regarding sterilizer use and sterilization monitoring. In addition, the Centers for Disease Control and Prevention (CDC), American Dental Association (ADA) and the Organization for Safety and Asepsis Procedures (OSAP) may be contacted for further information regarding infection control.
Page 20
Use an FDA approved biological indicator (BI) weekly to confirm sterility and STEAMPlus Class 5 integrators with each load. Follow the BI manufacturer's instructions. Revolutionary Science recommends the use of biological indicators and class 5 integrators manufactured by SPS Medical.
Page 21
Saniclave RS-SC-200 Proper placement of the door seal. R E V O L U T I O N A R Y S C I E N C E...