Saniclave RS-SC-200
5.
Low Water:
6.
Exhaust tank full:
7.
Door is difficult to open or close:
8.
Display reads '8888'
R E V O L U T I O N A R Y S C I E N C E
The display will read 'Err 5' if there is an insufficient
amount of water in the chamber. Press SET to
acknowledge the error. Fill reservoir up to the tempera
ture sensing probe. Restart cycle.
After every five cycles, the Saniclave display will flash
'drn tank full.' At this point, the exhaust tank must be
emptied if the end of the drain tube is not already rest -
ing in a sink or bucket. Revolutionary Science recom
mends that the drain tube be left in the sink or other
reservoir for regular drainage. The drain tank warning
can be bypassed by pressing any button. If exhaust
tank is not drained after every sixth cycle, water may
overflow.
There may be a few reasons why the door is difficult to
open or close.
1) Ensure that the door seal is tucked under all lugs in the
door. (see picture on page 19)
2) The gasket offset spring must rest against the outside
diameter of the seal and should not overlap the seal.
3) Check to make sure door markings line up properly.
Misaligned door markings will inhibit the door functionally
4) The chamber may have not yet depressurized.
Wait up to five minutes and then attempt to open the
door.
5) the door stop may be caught on the door clip near the
bottom of the door. This can be dislodged by pushing up
on the bottom of the plastic door housing.
6) The door seal may need lubricant. To do this, remove
and dry the seal completely. lightly dust the seal with
talcum powder and spread evenly around the entire seal
(or talc based baby powder) approximately once every 50
to 100 cycles or as needed. Caution: Too much
lubrication may compromise the seal.
This means that your real time clock has reset. The time
and date settings will reset if the Saniclave is left un
plugged longer than three days. See page 9 for setting the
date and time. Unplug and replug the power cord. The
display should now read '0000' and you can commence
with sterilization. The real time clock does not need to
be set properly in order to operate the Saniclave.