Saniclave Front-Loading Series INTRODUCTION Dear Valued Customer, Thank you for purchasing the Revolutionary Science Saniclave. We are confident that the Saniclave will provide years of reliable service. The high quality of construction and durable parts deliver a performance equivalent to much costlier equipment.
INTENDED USE The Revolutionary Science Saniclave is designed to be used in medical and dental clinics, hospitals and other facilities where reusable sterile equipment is used. It is intended to sterilize wrapped and unwrapped heat and moisture stable solid...
Saniclave Front-Loading Series LIMITATIONS OF USE Revolutionary Science is not responsible for misuse of the Saniclave. 1.The load should be considered non-sterile unless the display reads ‘CC’ after the cycle. 2. NEVER attempt to dismantle or modify the Saniclave. 3. NEVER leave an operating Saniclave unattended.
Saniclave Front-Loading Series OPERATING THE SANICLAVE PREPARATION FOR STERILIZATION (DEFAULT MODE) Important:The load should be considered non-sterile unless the display reads ‘CC’ after the cycle. 1. Preclean instruments according to their manufacturer’s recommendations. WARNING! Failure to properly clean instruments before autoclaving, may result in unsterile equipment.
Page 9
Saniclave Front-Loading Series OPERATING INSTRUCTIONS 1. Place the Saniclave on a dry, level surface and rest the drain tube in a sink, bucket, or other receptacle. Alternatively, the drain tube may be pinched shut and the water drained every few cycles (see pg. 10). 2.
Page 10
Saniclave Front-Loading Series 17. When the chamber has completed draining the autoclave will make an audible beep. and the screen will read ‘turn door’ or ‘CC’ after wrapped or unwrapped sterilization cycles respectively. If you autoclaved a wrapped load, rotate the door counterclockwise, remove the door from the chamber, and tuck the bottom lip of the door under the retainer clip so that the door gently rests against the autoclave providing a small air gap for the drying cycle.
Page 11
Saniclave Front-Loading Series SETTING YOUR SANICLAVE The sterilization cycle parameters are as follows: 30 minutes, 121 °C, 15 psi. These parameters cannot be adjusted. For wrapped loads, the user must use wraps or pouches that are cleared by the FDA for human use.
Saniclave Front-Loading Series TROUBLESHOOTING Please visit the Revolutionary Science troubleshooting webpage by scanning the QR code or entering it in manually to view instructional videos covering potential Saniclave issues and their solutions. https://www.revsci.com/pages/troubleshooting. 17319 Lake Blvd. Shafer, MN 55074 | 800-775-7942 | Fax: 775-218-7015 | support@revsci.com | revsci.com...
Page 13
Note: If a failure occurs, the load must be considered non-sterile. The sterilizer and load should then be investigated. Contact Revolutionary Science if problem persists. 1. Manual Cycle Abort: If ‘START’ is pressed, display will read ‘Err 1’ for cycle abort.
Page 14
Water Leakage: Water may leak from the unit if the drain tube is not placed in a sink or water receptacle. To troubleshoot first check the drain tube. If leaking comes from another source, unplug the machine and contact Revolutionary Science. If any of the above occurs, turn off the Saniclave and investigate. If the problem persists, contact Revolutionary Science.
WARRANTY Your Revolutionary Science Saniclave is warrantied to be free from defects of material and workmanship under normal use for a period of two years from date of purchase. If your Saniclave proves to be defective within this warranty period, APR&D, Inc.
ISO 9001: Revolutionary Science is an ISO Registered Company IQ/OQ/PQ: Revolutionary Science provides a range of compliance statements and IQ/OQ/ PQ documents that will enable customers to successfully validate their products in accordance with regulations and standards. Contact Revolutionary Science for IQ/OQ/PQ documentation.
______________________________________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________ Revolutionary Science recommends the weekly use of G. Bacillus Stearothermophilus biological indicator ampules. Place indicator in center of full load and test weekly. The use of FDA cleared class 5 integrators should be used with each load. Follow the manufacturer’s recommendations for class 5 integrator use.
All repairs must be performed by a trained professional. Contact Revolutionary Science for more information. (See the contact section on pg. 13) 17319 Lake Blvd. Shafer, MN 55074 | 800-775-7942 | Fax: 775-218-7015 | support@revsci.com | revsci.com...
Page 19
Saniclave Front-Loading Series SANICLAVE CYCLE DOCUMENTATION TABLE It is recommended to keep records of each sterilization cycle. This table may be copied. 17319 Lake Blvd. Shafer, MN 55074 | 800-775-7942 | Fax: 775-218-7015 | support@revsci.com | revsci.com...
Saniclave Front-Loading Series STERILIZATION MONITORING GUIDELINES Note Revolutionary Science provides the following information only for reference. The user must contact state and local government for proper regulations regarding sterilizer use and sterilization monitoring. Additionally, the Centers for Disease Control and...
Page 21
Use an FDA approved biological indicator (BI) weekly to confirm sterility and STEAMPlus Class 5 integrators with each load. Follow the BI manufacturer’s instructions. Revolutionary Science recommends the use of biological indicators and class 5 integrators (not supplied by Revolutionary Science).
Saniclave Front-Loading Series Door Seal Proper placement of the door seal. The door seal should be tucked under the lugs. 17319 Lake Blvd. Shafer, MN 55074 | 800-775-7942 | Fax: 775-218-7015 | support@revsci.com | revsci.com...
Page 23
Saniclave Front-Loading Series Rear of Saniclave Allow to the Saniclave to drain during each run by placing the drain tube in a receptacle or empty the exhaust tank via drain tube every few cycles. Exhaust Tank Drain Tube 17319 Lake Blvd. Shafer, MN 55074 | 800-775-7942 | Fax: 775-218-7015 | support@revsci.com | revsci.com...