Summary of Contents for Revolutionary Science Saniclave RS-SC-102

  • Page 2 Saniclave RS-SC-102 'Based on user feedback, we have supplied your Saniclave with two new design updates. New, high capacity basket. This new basket (not yet shown in the video or our website) has an increase of 1.75" in height and 0.4" in diameter over the original basket design.
  • Page 3: Table Of Contents

    Use an FDA approved biological indicator (BI) weekly to confirm sterility and STEAMPlus Class 5 integrators with each load. Follow the BI manufacturer's instruc- Warranty, Registration and Policy tions. Revolutionary Science recommends the use of biological indicators and class 5 Specifications integrators manufactured by SPS Medical.
  • Page 4: Introduction

    In addition, the Centers for Disease Control and Prevention (CDC), American Dental Association Thank you for purchasing the Revolutionary Science Saniclave. We are confident that (ADA) and the Organization for Safety and Asepsis Procedures (OSAP) may the Saniclave will provide years of reliable service.
  • Page 5: Display, Symbols, Control And Parts

    Saniclave RS-SC-102 Saniclave RS-SC-102 Saniclave Cycle Documentation Table Display, Symbols, Control and Parts It is recommended to keep records of each sterilization cycle. This table may be copied. R E V O L U T I O N A R Y S C I E N C E...
  • Page 6: Installation Instructions

    For regular Sterilizer Monitoring Guidelines, see page 16 Saniclave 102 —Indications For Use (applies to the Saniclave 102): The Revolutionary Science Saniclave is designed to be used in medical and dental clinics, hospitals and other facilities where reusable sterile equipment is used. It is intended to...
  • Page 7: Intended Use

    Hot water may drip from cover during removal and can cause scalding. Part: Model number: Limitations of Use: _ Autoclave 200-A00001 Revolutionary Science is not responsible for misuse of the Saniclave 1. NEVER attempt to dismantle or modify the Saniclave. _ Door 100-A00002 2. NEVER leave operating Saniclave unattended.
  • Page 8: Operating The Saniclave

    RS-SC-102 is a 510k listed product for medical use. K122978 result in unsterile equipment. Using unsterile equipment may result in se- FDA Registered: Revolutionary Science is an FDA Registered Company rious injury or death. CE / IEC Safety: RS-SC-102 tested to and found to be compliant with IEC 61010A 4.It is recommended that all instruments be dismantled, if possible.
  • Page 9 To view the cycle time left, press 'START.' This Your REVOLUTIONARY SCIENCE Saniclave is warranted to be free from defects of will not abort the cycle. It will change the reading on the display to cycle time.
  • Page 10 Restart cycle. If an over-temperature condition occurs again, con- Standby Mode Display reads “-00” tact Revolutionary Science to investigate the problem further. Discontinue Preheat Mode Display will read 'Ph' for Preheat. This mode will use until problem is resolved.