Extended Overhead Analysis Introduction Functions The “Extended Overhead Analysis” option provides further useful functions for your ANT-20: • Recording of selected Section Overhead (SDH) or Transport Overhead (SONET) bytes. • Recording of selected Low or High Path Overhead (POH) bytes.
ANT-20/ANT-20E Extended Overhead Analysis Operation Recording selected bytes (Overhead Capture) Function The “Capture” function is opened from the “Overhead Analyzer” window. After the byte has been selected using the cursor, the “Capture” dialog is opened using the Capture menu command.
ANT-20/ANT-20E Extended Overhead Analysis Recording results 1.3.1 Recording K1, K2 bytes, APS channel ⇒ Click the “START” button to prepare to start the recording. The record starts as soon as the pre-set trigger condition occurs. The changes in the selected byte are recorded frame by frame with timestamp.
Extended Overhead Analysis ANT-20/ANT-20E 1.3.2 Recording TCM frames ⇒ Click the “START” button to prepare for recording. Recording starts as soon as the TCM FAS word is detected. Complete TCM frames are captured and displayed. The first eight bytes, which contain the TCM FAS word, are also output.
ANT-20/ANT-20E Extended Overhead Analysis 1.3.3 Recording further SOH/TOH or POH bytes ⇒ Click the “START” button to prepare to start the recording. The record starts as soon as the pre-set trigger condition occurs. The changes in the selected byte are recorded frame by frame with timestamp.
The switching time must meet certain criteria. The ANT-20 checks these criteria by measuring the length of time that a certain event persists (e.g. AIS or loss of test pattern) after APS is triggered. The measured time is compared with a pre-set threshold value.
ANT-20/ANT-20E Extended Overhead Analysis 2.2.1 Setting the switch event The event to be used to trigger switching and to be measured is set from the “Sensor” list box. • MS-AIS measures the duration of a Multiplex Section Alarm. • TSE measures the duration of loss of test pattern (PRBS).
Page 15
ANT-20/ANT-20E Extended Overhead Analysis Measurement Settings 1. Click on the APS icon in the toolbar of the “Signal Structure” VI. The “APS Time Measurement” dialog opens. Fig. O-13 “APS Time Measurement” dialog 2. Select an event for triggering the switch operation from the “Sensor” list box.