Set Data Storage Criteria - Accumet XL 20 User Manual

Xl series meters
Table of Contents

Advertisement

This screen allows you to select which criteria are stored in the data logger with the measurement
when you store the data .The status of the current data storage criteria is displayed on the
screen. The criteria option is active if ON appears to the right of the option. It is inactive if OFF
appears to the right of the option. Any active criteria will be stored on demand.
To set Data Storage Criteria
1. Touch the Touch here to edit button next to the Data Storage Criteria to access the pH
Data Storage Setup screen.
2. Touch ON or OFF to change the status of a criteria you want to modify.
3. Repeat step 2 for all the remaining criteria except Data logging Interval (or Data storage
Interval).
4. For Data logging Interval (or Data storage Interval), touch MANUAL to store Ion data only
when the Log Data button is pushed, touch STABLE to automatically store Ion data when
Ion reading is stable, or touch TIMED to set a specific timed interval in seconds to store Ion
data.
5. Touch OK button to accept the changes of the entire group of data storage criteria and
return to the Ion Setup screen.
Ion Setup: Set Data Storage Criteria
XL25, 50 and 60 meters
59

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

XlXl 25Xl 1530Xl 60Xl 40 ... Show all

Table of Contents