Using the XL Series Meter: Button Functions
The following buttons appear in the XL series meters Mode screen:
*Replatinization is the process of replacing the platinum on the surfaces of the 2 cell
conductivity probes that may flake or wear off over time. The platinum on the surface of
the probe is used to increase the surface area of the measuring surface resulting in
decreased polarization error. Replatinization is a relatively quick procedure to perform,
taking no more than 5 minutes. See 'Appendix: Replatinization' for procedure.
*Note: Replatinization function is available in XL20, XL30, XL50 and XL60 meters only.
The About button lets you view the software revision and serial number details.
The Sys.Setup button allows you to turn on or off the beep status. This is the audible
beep that sounds each time any button is touched.
The *Multi Channel button lets you view multiple channels at the same time.
*Note: This is available in XL20, XL25, XL50 and XL60 meters only.
Single Channel Display
Dual Channel Display
Triple Channel Display
Quadraple Channel Display
The *Display Setup button lets you configure the display to read the inputs from the
channels of your choice.
*Note: Display Setup function is available in XL20, XL25, XL50 and XL60 meters only.
XL15, 20, 25, 30, 40, 50 and 60 meters
XL15 (Display - Ch1/Ch3)
XL30 (Display – Ch5)
XL40 (Display – Ch4)
XL20 (Display 1 & 2 - Ch1/Ch3/Ch5)
XL25 (Display 1 & 2 - Ch1/Ch2/Ch3)
XL50 (Display 1, 2 & 3 – Ch1/Ch2/Ch3/Ch5)
XL60 (Display 1 , 2 , 3 & 4 – Ch1/Ch2/Ch3/Ch4/Ch5)
23