126
c
6: Computer Interface Operation
HAPTER
IDN?
Input
Returned
Format
Example
OPC
Input
Remarks
OPC?
Returned
Remarks
RST
Input
Remarks
SRE
Input
Format
Remarks
Example
SRE?
Input
Returned
Format
Model 336 Temperature Controller
Identification Query
IDN?[term]
<manufacturer>,<model>,<instrument serial>/<option serial>,
<firmware version>[term]
s[4],s[8],s[7]/s[7],n.n
<manufacturer>
Manufacturer ID
<model>
Instrument model number
<instrument serial>
Instrument serial number
<option card serial>
Option card serial number
<firmware version>
Instrument firmware version
LSCI,MODEL336,1234567/1234567,1.0
Operation Complete Command
OPC[term]
Generates an Operation Complete event in the Event Status Register upon comple-
tion of all pending selected device operations. Send it as the last command in a com-
mand string.
Operation Complete Query
OPC?[term]
1[term]
Places a 1 in the controller output queue upon completion of all pending selected
device operations. Send as the last command in a command string.
Not the same as OPC.
Reset Instrument Command
RST[term]
Sets controller parameters to power-up settings.
Service Request Enable Register Command
SRE <bit weighting>[term]
nnn
Each bit has a bit weighting and represents the enable/disable mask of the corre-
sponding status flag bit in the Status Byte Register. To enable a status flag bit, send
the command *SRE with the sum of the bit weighting for each desired bit. Refer to
section 6.2.6 for a list of status flags.
To enable status flags 4, 5, and 7, send the command *SRE 208[term]. 208 is the sum
of the bit weighting for each bit.
Bit
Bit Weighting
4
16
5
64
7
128
Total:
208
Service Request Enable Register Query
SRE?[term]
<bit weighting>[term]
nnn (Refer to section 6.2.6 for a list of status flags)
Event Name
MAV
ESB
OSB