Null Detector; Versus Range Checks - Fluke 893A Instruction Manual

Voltmeter
Table of Contents

Advertisement

n.
Connect
the output
of
the
dc voltage calibrator
between
the
LO
and HI
INPUT
terminals.
o.
Apply
the dc input
voltages
versus
the
ranges
in-
dicated
in
Figure
4-5,
observing
that the
meter
indicates full-scale
(+1)
±
1 -
1/2
small
divisions.
4-30.
When
the
results
of
the
checks agree
with
the
information
given, the Null
Detector and Divider
cir-
cuitry
are
functioning
properly and
the test
equipment
can
be disconnected
from
the
instrument.
INPUT
DC VOLTAGE
RANGE
NULL
SENS
.001
1
1
mv
.01
1
.
Olv
.
1
1
.
1
v
1.0
1
TVM
10.0
10
TVM
100.0
100
TVM
1000.0
1000
TVM
Figure 4-5.
INPUT
DC
VOLTAGE
VERSUS
RANGE-CHECKS
4-31.
KELVIN-
VARLEY
DIVIDER
CHECKS
4-32.
The
Kelvin-
Varley Divider Checks require
the
use
of
a dc
voltage calibrator.
The
Precalibration and
Troubleshooting
Checks
of
paragraphs 4-21
through
4-30
must
also
have been
performed
to
ascertain
that
the
associated Instrument
circuitry
is
functioning
correctly.
To perform
the
checks, apply ac
power
to the
instrument and
proceed
as
follows:
a.
Place
the
Model
893 controls as follows:
POWER
RANGE
NULL
SENS
MODE
Readout
dials
ON/LTNE
OPR
10
.001
+
10.
000
-
A4TP11
FULL-SCALE INPUT
Ft"".
;
5 u
SATURATION
CONDITION
4
1
:ffz
5
u
—j
C
-
c
T...
u
A4TP13
A4TP15
Note:
HORIZONTAL
SWEEP
-
5
MSEC/CM
Figure
4-6.
NULL DETECTOR
WAVEFORMS
4-10
b.
Apply
+10
volts dc
from
a dc voltage calibrator
between
the
Model
893
LO
and HI
INPUT
terminals.
c.
Adjust
A2R33
for a
zero-center
meter
indication
on
the
Model
893
d.
Apply
the
dc
input voltages
versus
the
Readout
dial
settings
of
Figure
4-7,
observing
that
the
Model
893
meter
indicates
zero-center
within the
major
division deviation.
INPUT
DC
VOLTAGE
READOUT
DIAL SETTINGS
MAJOR
DIVISION
DEVIATION
11.000
10.
99
X
±5.5
10.
000
10.00
0
null
reference
10.
00
9-99
X
±5. 0
9.999
9.
99 90
±4. 99
8.
888
8.
88
80
±4,
44
7.777
'^77 20
±3. 88
6.666
6.66
^
±3.33
5.555
5.55
5(}
±2.77
4.444
4.44 40
±2.
22
3,333
3-33 30
±1,
66
2.222
2,22 20
±1.
n
1.111
1.
11
10
±
.55
1.000
1.00
Q
±
.5
Figure 4-7.
INPUT
VOLTAGES
VERSUS
READOUT
SETTINGS
-
CHECKS
4-33.
When
the results
of
these
checks agree
with the
information
given,
the Kelvin-
Varley
Divider
circuitry
is
functioning
properly and
the test
equipment
can be
disconnected
from
the instrument.
4-34.
RECORDER OUTPUT
OPTION CHECKS
4-35.
The Recorder
Output Option
Checks
require
the
use
of
a dc
differential
voltmeter.
Circuit
wave-
forms
are
illustrated
in
Figure
4-8,
To perform
the
checks, apply ac
power
to
the
instrument and proceed
as
follows;
a.
Connect
the
inputs
of
a dc
differential
voltmeter
to
the
rear panel HI and
LO
RECORDER OUTPUT
terminals and
rotate the
associated
level control
fully
clockwise.
b.
Connect
a
shorting
jumper between
the
LO
and HI
INPUT
terminals
of the
Model
893
and
place the
controls as follows:
POWER
RANGE
NULL
SENS
MODE
Readout
dials
ON/LINE
OPR
1
.001
+
.00100
c.
A
dc voltage
of -0. 2 to -0.
28
volts dc
should be
indicated
on
the
dc
differential
voltmeter connected
to
the
RECORDER OUTPUT
terminals.
893A

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

893ac

Table of Contents