Meter Specifications

Result Range: 1.1 - 33.3 mmol/L (20 - 600 mg/dL)
Sample Size: Minimum 1 microliter (1 µl)
Sample: Fresh capillary whole blood or Glucose Control
Test Time: 10 seconds
Result Value: Plasma values
Assay Method: Electrochemical
Power Supply: One 3V
Battery Life: Approximately 1,100 tests or 1 year
Automatic shut-off: After two minutes of non-use
Weight: 47 grams (1.66 ounces)
Size: 89 x 55 x 17mm (3.52" x 2.15" x 0.67")
Memory Size: 200 blood glucose results,
System Operating Range
Relative Humidity: 10-90% (Non-condensing)
Temperature: 10º- 40ºC (50º-104ºF)
Hematocrit: 30-55%
Note: Use within specified environmental
conditions only.
#CR2032 (non-rechargable)
Total power when active
at full battery = 8.6mW
1 Control Test result
29
Lithium Battery
(Meter & Test Strips):