Test Strip Care
• Store Strips in original vial only.
• Write date opened on Strip vial. Discard unused
Strips from vial 120 days after opening or after
date printed next to symbol, whichever comes
first. Use of Strips past either date may give
incorrect results.
• Close vial immediately after removing Strip. Use
Strip quickly after removal from vial. Never store
Strips outside of vial. Store in a dry place at
room temperature below 86°F/30°C. DO NOT
FREEZE.
• Do not reuse Strip.
• Apply only fresh capillary blood or TrackEASE
Glucose Control to Strips.
• Do not bend, cut or alter Strips in any way.
Code Chip Care
• Code Chip must remain in Meter until all Strips
from box are used, then discard old Code Chip.
• Insert new Code Chip from new box of Strips
immediately before using Test Strips.