Quality Control Tests; Check Strip Test - Invacare trackease smart system User Manual

Table of Contents

Advertisement

Available languages

Available languages

Quality Control Tests

Check Strip Test

Check Strip is stored in Carrying Case. Check
Strip Test is used to check that Meter is working
properly. Perform a Check Strip Test when:
• Using Meter for first time.
• Results do not match how you feel.
• Meter was dropped or banged.
1. Start with Meter off. Remove
Check Strip from Carrying Case.
Insert Check Strip Contact End Œ
into Test Port with dimple facing
up. Meter turns on.
2. Check Display to make sure that
all segments are showing. (See
Pg. 3.)
3. After a few seconds, Meter
displays
Caution! If segments are missing from Display
4. Remove Check Strip. Return to Carrying Case.
Meter turns off.
if test is correct.
or
does not appear, do not use
Meter for testing. Call for assistance.