Mounting Position; Mounting; Torques And Pressures - Honeywell VERSAFLOW MAG 1000 Quick Start Manual

Electromagnetic flow sensor
Table of Contents

Advertisement

VERSAFLOW MAG 1000

2.8.10 Mounting position

Figure 2-18: Mounting position

2.9 Mounting

2.9.1 Torques and pressures

The maximum pressure and torques values for the flowmeter are theoretical and calculated for
optimum conditions and use with carbon steel flanges.
Figure 2-19: Tightening of bolts
Tightening of bolts
• Always tighten the bolts uniformely and in diagonally opposite sequence.
• Do not exceed the maximum torque value.
• Step 1: Apply approx. 50% of max. torque given in table.
• Step 2: Apply approx. 80% of max. torque given in table.
• Step 3: Apply 100% of max. torque given in table.
34-VF-25-54 iss.3 GLO July 14 US