b. Process the acquired data. Press "RES - RES0"
c. Press "OK" according to the message dialog box.
d. Press "Validation & Initialization & Defrag", and press
"OK" according to the message dialog box.
*When the setting of 25-point calibration is not
conformed, error message appears. Then follow
"To start performing 25-point calibration" again.
e. Press "DATA - Export" to save result of the calibration.
• The data will be saved as "importf.prm" in "25P
Adjustment Data Processing" folder.
• Press "Exit" to quit.
• Enter product's serial number when asked for S/N.
• "S/N.xls" will be saved in "25P Adjustment Data Pro-
cessing" folder.
3. Import the acquired data.
a. Press "Parameter" in LTabMonS.
b. Press "Load File" to open "importf.prm" file saved in
"25P Adjustment Data Processing" folder.