787
Users Manual
DC Millivolts Measurement
Range (mV dc)
400.0
AC Volts Measurement
Range (ac)
400.0 mV
4.000 V
40.00 V
400.0 V
1000 V
Specifications are valid from 5 % to 100 % of amplitude range.
AC conversion: true rms
Maximum crest factor: 3
For non-sinusoidal waveforms, add
Ω
Input impedance: 10 M
(nominal), < 100 pF, ac-coupled
Common mode rejection ratio: >60 dB at dc, 50 Hz, or 60 Hz
38
info@Fluke-Direct.com
Resolution
0.1 mV
Resolution
50 Hz to 60 Hz
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
±
(2 % reading + 2 % f.s.) typical
Fluke-Direct
Accuracy (% of Reading + Counts)
0.1 % + 1
Accuracy, ±(% of Reading + Counts)
45 Hz to 200 Hz
0.7 % + 4
1.2 % + 4
0.7 % + 2
1.2 % + 4
0.7 % + 2
1.2 % + 4
0.7 % + 2
1.2 % + 4
0.7 % + 2
1.2 % + 4
.com
1.888.475.5235
200 Hz to 500 Hz
7.0 % + 4
7.0 % + 4
7.0 % + 4
7.0 % + 4
7.0 % + 4