Filters:
Safety System:
Width:
Height:
Depth:
Weight:
Shipping Weight:
Operating Ambient Temperature and Humidity:
Cabinet:
Standards and Regulatory Listing:
Electrical:
Placement:
1118353-M-00
Cabinet (2), Outlet HEPA and Compressor Inlet
Current overload or line surge shutdown.
High temperature compressor shutdown.
High Pressure Alarm w/compressor shutdown.
Low Pressure Alarm w/compressor shutdown.
Battery Free Power Loss Alarm.
18 3/8 in ± 3/8 in (46.7 cm ± 1 cm)
26 3/8 in ± 3/8 in (67.0 cm ± 1 cm)
14 3/8 in ± 3/8 in (36.5 cm ± 1 cm)
53 lbs ± 2 lbs (24.0 kg ± 1 kg)
60 lbs ± 2 lbs (27.2 kg ± 1 kg)
50°F to 95°F (10°C to 35°C) at up to 60% relative humidity
Impact Resistant flame-retardant plastic cabinet that conforms to UL
94V–0
Double Insulated Product ETL certified complying with UL 1097 ETL
certified to CSA C22.2 No. 68.
No extension cords
No closer than 12 in (30.5 cm ) from any wall, furniture, draperies or
similar surfaces to assure sufficient air flow.
Avoid deep pile carpets and heaters, radiators or hot air registers.
Floor location only.
No confined spaces (Example: No closets).
Technical data
29