Invacare® Platinum™10L Oxygen Concentrator

6.3 Setting Up the Concentrator

1.
Plug in power cord to an electrical outlet.
2.
Connect Humidifier (if so prescribed).

6.3.1 Setting Up the Humidifier

WARNING!
Risk of Injury or Damage
To avoid burns from steam or hot water, inhalation of
water and/or water damage to the concentrator:
– DO NOT fill humidifier bottle with hot water. Allow
boiled water to cool to room temperature before
filling.
– DO NOT overfill humidifier.
– Replace the humidifier cap and securely tighten.
Confirm that the cap is not cross-threaded on the
humidifier bottle.
– DO NOT reverse the oxygen input and output
connections. Water from the humidifier bottle will
travel through the cannula back to the patient if input
and output connections are reversed.
– When using tubing connections longer than 7 ft
(2.1 m) in length, position the humidifier as close
to the patient as possible to allow for maximum
humidification output.
18
WARNING!
Risk of Injury or Damage
Failure to properly install the humidifier bottle or other
accessories to the concentrator will impact the flow of
oxygen.
– To check for proper operation of the oxygen
concentrator and accessories, place the end of the
nasal cannula under the surface of a half-full cup of
water only when the oxygen is flowing and look for
bubbles indicating proper operation. Wipe any excess
water off the nasal prongs when finished before placing
on the patient.
– If there are no bubbles, check all connections
(including humidifier bottle and other accessories, if
applicable) and repeat. Contact your dealer or service
supplier immediately if bubbles still do not appear.