System fails promptly after showing brand logo
System fails with continuous BEEP sound and erratic
display
Nothing different.
It stops at the logo display
Report to your local distributor
Internal 1MB memory of Scanner seems defective
and needs reflashing.
Replacement of the main board may be
necessary, too.
Report the symptom to your local distributor and
wait for the instruction.
Remove power then turn ON again
No, it doesn't work. It stops again.
Remove power.
Remove the cartridge from the slot
then Turn power ON.
insert the other cartridge and then
It shows the same trouble.
The system stops shortly after
showing the logo display
Chapter 7 - 8
For MULTISCAN P2
"ERROR!! INVALID NAND FLASH"
Error message appears and
key sounds when pressed
Turn the power off,
Restart
It works O.K. with the other cartridge
Cartridge is defective
The software cartridge seems defective.
Let your local distributor format the defective
cartridge and then rewrite the software for you.