Table 2 9103 Controller Communications Commands continued
Command Description
Read R0 calibration parameter
Set R0 calibration parameter to n r[0]=n
Read ALPHA calibration
parameter
Set ALPHA calibration parameter
to n
Read DELTA calibration
parameter
Set DELTA calibration parameter de[lta]=n
Read BETA calibration parameter be[ta]
Set BETA calibration parameter
Functions not on menu
Read firmware version number
Read structure of all commands
Read all operating parameters
Legend:
Note:
info@Fluke-Direct.com
Command
Command
Format
Example
r[0]
r
r=100.324
al[pha]
al
al[pha]=n
al=0.0038433
de[lta]
de
de=1.3742
be
be[ta]=n
be=0.342
*ver[sion]
*ver
h[elp]
h
all
all
[] Optional Command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a
carriage return and linefeed. Then the value is returned as indicated in the RETURNED column.
Fluke-Direct
8 Digital Communication Interface
Returned
Returned
Example
r0: 999.999
r0: 100.578
al: 9.9999999
al: 0.0038573
de:9.99999
de: 1.507
be:9.999
be:0.342
ver.9999,9.99
ver.9103,2.00
list of commands
list of parameters
.com
Interface Commands
Acceptable
Values
90 to 110
.002 to .005
0–3.0
–100.0 to 100.0
35
1.888.475.5235