6. CALIBRATION
6.1 STANDARD CALIBRATION
- Turn off the power, empty the pan, hold the ON/OFF key and the TARE key together until, display
will show "88888", now release the ON/OFF key;
- When the display will shows "F1 CAL", release also the TARE key;
- Press ZERO key to enter setting, display will show "UNLD" and after the instrument will acquired
automatically the zero value, after that it will show "LD";
-Put the same weight of the maximum capacity on the scale and wait the stability, the display will
show "PASS"
- Press ZERO key to return in the setup environment.
10