Physical Specifications Of The Agilent 2100 Bioanalyzer - Agilent Technologies 2100 Installation And Safety Manual

Bioanalyzer
Table of Contents

Advertisement

2
Site Requirements for the Agilent 2100 Bioanalyzer

Physical Specifications of the Agilent 2100 Bioanalyzer

Physical Specifications of the Agilent 2100 Bioanalyzer

Table 1
Bioanalyzer.
Table 1
Physical specifications
Type
Weight
Dimensions
Line voltage
Line frequency
Power consumption
Ambient operating temperature
Ambient non-operating
temperature
Humidity
Operating altitude
Non-operating altitude
Safety standards: IEC, CSA, UL
20
on page 20 lists the physical specifications of the Agilent 2100
Specification
10 kg (22 lbs)
162 — 412 — 290 mm
(6.4 — 16.2 — 11.4 in)
100-240 VAC ±10%
50-60 Hz ±5%
30 W / 60 VA
5-40 °C (41-104 °F)
-40-70 °C (-40-158 °F)
<80%, at 5-31 °C (41-89 °F), decreasing
linearly to 50% at 40 °C (104 °F)
Up to 2000 m (6500 ft)
Up to 4600 m (14950 ft)
Installation Category II, Pollution
Degree 2
Comment
Width — depth — height
Wide-ranging capability
Maximum
The assays set-up to be analyzed on a chip
temperature of 30 °C. If the ambient
temperature is above 30 °C the assays are
analyzed using that temperature.
Non-condensing
For storage of the Agilent 2100 Bioanalyzer

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents