Recommended Test Schedule

The schedule shown in
performance testing for a microplate reader used two to five days a week.
Note: The risk factors associated with your tests may require
that the Performance and Operational Qualification procedures
be performed more or less frequently than shown below.
System Self-Test, p. 107
Checksum Test, p. 109
Absorbance Plate Test,
p. 110
Liquid Test 1, p. 119
Liquid Test 2, p.121
Liquid Test 3, p. 125
Optional for 340 nm
*Run Liquid Test 2
If you run Liquid Test 2, you do
ELx800 Operator's Manual
Table 1
defines the factory-recommended intervals for
Table 1

Recommended Test Schedule

Installation
Performance
Qualification
Qualification:
Monthly
ONLY
if you do not have an Absorbance Test Plate.
Recommended Test Schedule |
Performance
Qualification:
Semiannually
*
NOT
have to also run Liquid Test 1.
105
Operational
Qualification:
Initially and
Annually
*