Data Logging

The Data Logger function makes the multimeter a programmable tool for
taking and storing a series of timed measurement readings. This feature
can be used in combination with any primary measurement function. The
readings are stored in non–volatile memory, and can be accessed from the
front panel or remote interface.
To customize a data logging event, the multimeter allows you to set the
following parameters: START DELAY, INTERVAL, TIME/EVENT (DURATION or COUNT).
• START DELAY is the delay time, settable from 0 to 3600 seconds, from
• INTERVAL is the time, settable from 0 to 3600 seconds, between
• TIME/EVENT is the total duration of the data logging event. You may
• Front Panel Operation: To set up a data logging event, press
34410A/11A User's Guide
initiating a trigger before the data logging function takes its first
reading.
subsequent readings.
choose to set a DURATION in hours, minutes and seconds, from 1 second
to 100 hours, or a COUNT of total readings, settable from 1 to 50,000
readings (model 34410A) or up to 1 million readings (model 34411A).
display DATA LOGGER. Select SETUP and press
Use the navigation keypad to set the desired start delay, and press
to display INTERVAL. Use the navigation keypad to set the desired reading
interval, and press
needed to select DURATION or COUNT, and press
If DURATION was selected, use the navigation keypad to set the desired
reading interval, and press
navigation keypad to set the desired total number of readings, and
.
press
Upon pressing
TRIGGER KEY. The data logger is now programmed. The primary display
line will then show the selected primary measurement function. The
second display line will show a progress bar and the count of set
total readings. These displays will remain static until you press
or pulse the rear panel Ext Trig terminal.
When you press
after the set start delay has expired. While it is reading, the primary
display line will show each reading, and the second display line will
show a progress bar and a countdown of readings remaining. The
multimeter will continue taking readings until the programmed total
readings count has been reached, or unless the data logger is stopped
manually.
to display TIME/EVENT. Press
. If COUNT was selected, use the
, the displays will briefly read TO START and PRESS
, the multimeter will begin to take readings,
Features and Functions