• Remote Interface Operation: The following commands set the aperture
• The following table indicates the resulting resolution for each
AC Measurements
See
current and frequency/period measurements.
34410A/11A User's Guide
value, in seconds:
[SENSe:]VOLTage[:DC]:APERture {<seconds>>MIN>MAX>DEF}
[SENSe:]CURRent[:DC]:APERture {<seconds>>MIN>MAX>DEF}
[SENSe:]RESistance:APERture {<seconds>>MIN>MAX>DEF}
[SENSe:]FRESistance:APERture {<seconds>>MIN>MAX>DEF}
[SENSe:]TEMPerature:APERture {<seconds>>MIN>MAX>DEF}
The following command (similar queries for current, resistance and
temperature) queries the aperture setting:
[SENSe:]VOLTage[:DC]:APERture?[{MIN>MAX}]
The following command (similar queries for current, resistance and
temperature) queries if aperture mode is enabled:
[SENSe:]VOLTage[:DC]:APERture:ENABled?
This query returns a "0" (disabled) or "1" (enabled)
Refer to the Agilent 34410A/34411A Programmer's Reference for
complete description and syntax for these commands.
integration time value (in power–line cycles) selected.
Integration Time
0.001 PLC
0.002 PLC
0.006 PLC
0.02 PLC
0.06 PLC
0.2 PLC
1 PLC (default)
2 PLC
10 PLC
100 PLC
*
These two settings are
provided on model 34411A only
"AC Filter"
on page 56 for configuration options for ac voltage, ac
*
*
.
Features and Functions
Resolution
30 ppm x Range
15 ppm x Range
6.0 ppm x Range
3.0 ppm x Range
1.5 ppm x Range
0.7 ppm x Range
0.3 ppm x Range
0.2 ppm x Range
0.1 ppm x Range
0.03 ppm x Range