Ranging - Agilent Technologies 34410A User Manual

6 1/2 digit multimeter
Hide thumbs Also See for 34410A:
Table of Contents

Advertisement

Ranging

This feature applies to voltage, current, resistance, capacitance and
frequency/period measurements only.
You can let the multimeter automatically select the range using
autoranging, or you can select a fixed range using manual ranging.
Autoranging is convenient because the multimeter automatically selects the
appropriate range for sensing and displaying each measurement. However,
manual ranging results in better performance, since the multimeter does
not have to determine which range to use for each measurement.
• Autorange thresholds – The multimeter will shift range as follows:
• Ranging is local to the selected function; the multimeter remembers the
• For frequency and period measurements, the selected range is for the
• For temperature measurements, the multimeter uses auto ranging,
• For continuity tests, the multimeter range is fixed at 1 kW, and for
• Front Panel Operation: For applicable measurement functions, the ranging
34410A/11A User's Guide
Down range at <10% of current range
Up range at >120% of current range
selected ranging method (auto or manual) and selected manual range
when you switch between measurement functions.
ac voltage level of the input signal.
depending on the probe type.
diode tests the range is fixed at 1 VDC with a 1 mA current source
output.
method and manual range may be selected by front panel shortcuts (see

"Ranging"

on page 43), or through the configuration menu for the
selected function. For the latter method, press
displayed, then select AUTO or MANUAL using the navigation keypad and
. If MANUAL has been selected, press
press
select the desired range, and press
configuration menu.
Manual ranging – If the input signal is greater than can be measured on
the selected range, the multimeter provides these overload indications:
\
OVLD from the front panel or "
Features and Functions
or
. Then step through or exit the
\
9.9E+37" from the remote interface.
2
until RANGE is
as needed to
59

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

34411a

Table of Contents