Required Test Equipment - HP 3400A Operating And Service Manual

Rms voltmeter
Hide thumbs Also See for 3400A:
Table of Contents

Advertisement

r
Section
V
Model 3400A
Table
5-1.
Required Test Equipment
INSTRUMENT
TYPE
REQUIRED CHARACTERISTICS
USE
RECOMMENDED MODEL
DC
Voltmeter
Accuracy:
±0.
1%
full
scale
Voltage Range:
lOmVto
100
V
Performance Checks
Adjustment and
Calibration
-hp-
Model 3440A/3443A
Digital
Voltmeter
Voltmeter
Calibrator
Voltage
Range:
1
mV
to
300
V
rms
Frequency: 400
Hz
Performance Checks
Adjustment and
Calibration
Troubleshooting
-hp-
Model 738B
Volt-
meter
Calibrator
Oscillator
Frequency Range:
10
Hz
to 10
MHz
Output:
1
mV
to
3
V
Frequency Response: 0.25%
(expand
scale)
Performance Checks
Adjustment and
Calibration
Troubleshooting
-hp-
Model 652A
Test
Oscillator
Oscilloscope
Sensitivity:
0.
005
v/cm
Bandwidth:
dc
to
20
MHz
Adjustment and
Calibration
T
rouble sho
ot
ing
-hp-
Model 140A/
1402A/1420A
Pulse
Generator
Pulse
Width: variable
to
10
/isec
Pulse
Amp:
±10
volts peak,
variable
Pulse
Rate:
250
to
1000 pps
Performance Checks
-hp-
Model 214A
Pulse Generator
Frequency
Counter
Range: 250
to
1000
Hz
Accuracy:
±1
count
Time
Interval:
1
/isec
Performance Checks
-hp-
Model
5233
L
Electronic Counter
Peak
Responding
Voltmeter
Voltage Range:
0.
5
V
to 300
V
Accuracy:
±3%
full
scale
Performance Checks
-hp-
Model 410C
Voltmeter
Average
Responding
Voltmeter
Voltage
Range:
0.
001
to
300
V
Accuracy.
1%
full
scale
Adjustment and
Calibration
Troubleshooting
-hp-
Model
400E/EL
Voltmeter
Current
Supply/
Ohmmeter
Output:
5
mA
of
current
into
100ft
Ohm
Range:
1ft
to 10
Mft
Accuracy:
±5%
Troubleshooting
-hp-
Model 412A
DC
Vacuum
Tube
Voltmeter
RMS
Respond-
ing
Volt-
meter
Range:
1
mV
full
scale
Performance Checks
-hp-
Model 3400A
RMS
Voltmeter
Resistor
200
kft,
metal
film,
1/4
W
1%
Performance Checks
-hp-
Part No. 0757-0782
Resistor
499
kft,
metal
film,
1/4
W
1%
Troubleshooting
-hp- Part No.
0757-0327
Capacitor
1
iiF
T
rouble shooting
-hp- Part No.
0180-0269
50
ft
Feedthru
Termination
Resistor: fixed corap
50ft
±5%
1/4
W
Performance Checks
-hp-
Model 11048B
50ft
Feedthru
BNC-T-
Adapter
Performance Checks
Adjustment and
Calibration
-hp- Part No. 1250-0072
Adapter
4
10C
to
Dual
Banana
Performance Checks
-hp-
Model 11018A
Extender
Board
15 pin
programmable
Troubleshooting
-hp- Part No. 5060-6038
I
5-0
t

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents