Collector Of Q602; Base Of Q604 - HP 3400A Operating And Service Manual

Rms voltmeter
Hide thumbs Also See for 3400A:
Table of Contents

Advertisement

Section
V
Model 3400A
%
i
COLLECTOR OF
Q602
Insert
+
10
mV
signal
through
499
kn
resistortopinl3onA6
Sweep
=
5
ms/cm
Vertical
=
0.5
V/cm
DC
coupled
use
10:1
probe
Figure
5-10.
Collector
of
Q602
(feedback loop open)
Table
5-6.
Troubleshooting Procedure
(Cont'd)
CHECK
PROCEDURE
ACTION
13
Measure
the ac signal
at
base
of
Q604.
Refer
to
Figure 5-11
for
waveform.
PASSES: Proceed
to
@
FAILS:
Proceed
to
©
V
^*
L
nn
JU
BASE
OF
Q604
Insert +
10
mV
signal
through
499
kft
resistor
to pin 13
on
A6
Sweep
=
5
ms/cm
Vertical =
5V/cm
AC
coupled
no probe required
:
Figure
5-11.
Base
of
Q604
(feedback loop open)
'
5-12
01768-4