Intrinsically Safe electrical parameters

Table 1: Device Communicator communication module
Ui
Ii
Pi
Ci
Li
Uo
Io
Po
Co
Lo
Table 2: Device Communicator Plus communication module
mA
interface
mA
Ui
30 Vdc
Ii
200 mA
Pi
1.0 W
Ci
0
Li
0
U
0
o
Io
0
P
0
o
C
o
L
o
Quick Start Guide
FOUNDATION
fieldbus
(non-FISCO)
FF + and -
30 Vdc
380 mA
1.3 W
0
0
1.89 V
32 µA
61 µW
14.3 µF
100 mH
FOUNDATION
fieldbus
(non-FISCO)
FF pwr
and F-
FF + and -
17.5 Vdc
380 mA
1.3 W
231 nF
0
17.31 V
199 mA
0.94 W
-
See table
3
-
See table
3
FOUNDATION
(FISCO)
FF + and -
30 Vdc
215 mA (IIC) 380 mA (IIB)
1.9 W (IIC) 5.3 W (IIB)
0
0
1.89 V
32 µA
61 µW
14.3 µF
100 mH
HART
HART +
pwr
30 Vdc
30 Vdc
380 mA
200 mA
1.3 W
1.0 W
0
0
0
0
1.89 V
25.69 V
32 µA
105 mA
61 µW
668 mW
14.3 µF
See table
4
100 mH
See table
4
fieldbus
HART + and -
FOUNDATION
®
(FISCO)
HART +
FF pwr
and -
and F -
30 Vdc
17.5 Vdc
200 mA
380 mA
1.0 W
1.3 W
0
231 nF
0
0
1.89 V
17.31 V
1.9 mA
199 mA
3.6 mW
0.94 W
14.3 µF
See table 3
100 mH
See table 3
®
HART
30 Vdc
200 mA
1.0 W
0
0
1.89 V
32 µA
61 µW
14.3 µF
100 mH
fieldbus
FF + and -
30 Vdc
215 mA (IIC)
380 mA (IIB)
1.9 W (IIC)
5.3 W (IIB)
0
0
1.89 V
32 µA
61 µW
14.3 µF
100 mH
13