CPR Dashboard with Real CPR Help
See-Thru CPR
See-Thru CPR is a proprietary algorithm that filters the
CPR artifact from the ECG to allow users to see a close
approximation of a patient's organized underlying rhythm
during CPR compressions, reducing interruptions in CPR.
Activating See-Thru CPR:
1. Press Options softkey.
2. Press Traces softkey.
Using See-Thru CPR:
1. During CPR, view the filtered ECG
(second trace) on display to determine
if an organized rhythm appears to be
developing.
2. If rhythm appears to be organized
and shockable, press CHARGE while
continuing compressions.

CPR

3. Press Trace 2 softkey.
4. Press Filt ECG softkey.
3. Pause compressions and visually
confirm a shockable rhythm.
4. Shout "Clear" and verify others are clear.
5. Press SHOCK button.
6. Resume CPR immediately for 2 minutes
before stopping CPR to check rhythm. If
non-shockable, resume compressions.
Release Bar
Rate and Depth Indicators
Unfiltered ECG signal during CPR
Signal filtered by See-Thru CPR
PPI