Amptd; How To Set The Amplitude Output - HP 8642A Operating Manual

Synthesized signal generator
Hide thumbs Also See for 8642A:
Table of Contents

Advertisement

Operation
Amplitude
(cont'd)
Model 8642A/B
Procedures
Note:
Put the instrument in its initialized state by selecting 1 INSTR PRESET "1.
AMPTD
How to Set the Amplitude Output (with the EMF Mode Off)
You can set the amplitude output by selecting the amplitude function and then setting
the desired output level:
I AMPTD I and the desired output level (in dBm, V, mV, uV, or dBuV; the I RAD % I key
is the dBuV terminator).
Example
Set the RF output to -107.3 dBm.
LOCAL
(keys)
<aa^
(program
codes)
ENTRY
[AMPTD]
DATA
EKDHEDmi]
AP - 1 0 7 . 3 DM
Display
i iu kj.fcj t.i L I u u u MI c
. ' r a ~i 73
TJ
fn
I KJ
I. J
XI Ml
Amplitude has been set to -107.3 dBm. The cursor above the amplitude display setting
indicates amplitude is the active function.
Example
Set the RF output to 6 dBuV.
LOCAL
(keys)
4aan^
(program
codes)
ENTRY
[AMPTDJ
DATA
LsJ IT)
AP 6 DU
Display
, ' 0 0.0 0 0 0 0 0 CT Z
5.0 U 3
AMPTD
REL
Amplitude has been set to 6.0 dBuV. The AMPTD REL annunciators indicate that dBuV
is a relative setting. Even though dBuV can be set directly from the front panel, the way
this function operates and the way it is displayed are similar to most relative settings.
(Refer to Relative Amplitude, page 3-77.)
3-70
rev.20MAR85

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8642b8642a: 8642b

Table of Contents