Performance Tests
Model 8642A/B
PERFORMANCE TESTS
NOTE
This test is a good overall check of the phase noise
performance
of the HP 8642.
However,
if it is
necessary to check phase noise at a specific
output
frequency,
perform
test 4-16 or use an HP Phase
Noise Measurment
System to verify the
performance.
EQUIPMENT:
Mixer
HP 10514A
1 MHz Low Pass Filter
(See Figure 1-1)
Low-Noise Amplifier
HP 0 8 6 4 0 - 6 0 5 0 6
W/Power Supply HP 6216A
Spectrum Analyzer
HP 8566A/B
Oscilloscope
HP 19 80B
PROCEDURE:
Connect the instruments as shown in Figure 4 - 1 1 .
Preset the
instruments. Insure that the spectrum analyzer center frequency
calibration is accruate.
On the HP 8642, enter SPCL 8. Set the HP 8642 to 10.020 MHz and
- 5 0 dBm amplitude.
Set up the Spectrum Analyzer as follows:
VIDEO BW
SWEEP TIME
ATTENUATION
RESOLUTION
CENTER FREQ.
SPAN
REFERENCE LEVEL
1Hz
5.0 sec
OdB
10 Hz
20 kHz
0 Hz
- 5 0 dBm
NOTE
It may be necessary
to fine tune the spectrum
analyzer
center
frequency
adjustment.
The spectrum
analyzer
band width will be centered when the maximum
signal
level is obtained
while tuning
the
center
frequency
adjustment.
Record the level of the 20 kHz peak signal for later use.