Modulation Frequency Correction Off; Disable Settling; Calibrate The Modulation Frequency Bands Again - HP 8642A Operating Manual

Synthesized signal generator
Hide thumbs Also See for 8642A:
Table of Contents

Advertisement

Model 8 6 4 2 A / B
Operation
Modulation, $M (cont'd)
113.
Modulation
Frequency
Correction Off.
This special function
disables the
single-band calibration of the modulation frequency oscillator that normally occurs
whenever the modulation oscillator frequency is set or the internal modulation oscillator
is put into use. When this calibration is disabled, the typical modulation frequency
switching time is decreased by approximately 200 ms. For more information, refer to
"Speeding Up Modulation Switching Time", page 3-1 32.
213.
Modulation Frequency Correction On.
This special function
re-enables the
modulation frequency correction disabled with Special Function 11 3.
114.
Negative <t > M Polarity. This special function causes a positive voltage applied to
the external F M / ^ M INPUT to cause a phase lag and a negative voltage applied to cause
a phase lead.
214.
Positive $ M Polarity. This special function causes a positive voltage applied to
the external FM/<f>M INPUT to cause a phase lead and a negative voltage applied to
cause a phase lag.
119.
Disable
Settling.
This
special
function
allows
the
Signal
Generator's
microprocessor to respond to new commands without waiting for the internal circuitry to
settle. When using this special function, it's possible to select a function before the
previous function has finished settling.
When settling is disabled, response time is
improved by up to 40 ms. Special Function 119 disables out-of-lock messages.
219.
Re-enable Settling. This special function re-enables the settling that is disabled
when Special Function 119 is selected.
248. Calibrate the Modulation Frequency Bands Again. This special function enables
you to initiate the calibration sequence that normally only occurs during instrument
power-up.
This function can be used, for example, to recalibrate the modulation
oscillator frequency bands if normal modulation frequency correction has been disabled
with Special Function 11 3. Calibration typically takes one to two seconds.
Related
Operating
Instructions
Refer to Modulation Oscillator (page 3-131) for more information on the internal
modulation oscillator that is the common source for all internal modulation as well as for
the MOD OUTPUT connector.
Refer to Knob (page 3-141) and Step, Increment Set (page 3-149) for information on
using these to modify settings.
Refer to Operating Messages (page 3-177) if the instrument displays a message you want
clarified.
Refer to Special Functions (page 3-165) for information about the way these are
accessed.