HP 8642A Operating Manual page 275

Synthesized signal generator
Hide thumbs Also See for 8642A:
Table of Contents

Advertisement

Model 8642A/B
Performance Tests
PERFORMANCE TESTS
OSCILLOSCOPE
D O
□□
DDDDD □ n
cm
n
□ □ □ BnnD"
~
,-, r,
an D D D D
O O D D D
a n
B o n o
O
RF
OUTPUT R
HP 8642A/B
SYNTHESIZED
SIGNAL GENERATOR
MIXER
RF
OUTPUT
1 MHz
LPF
SPECTRUM
ANALYZER
INPUT
SIGNAL
GENERATOR
OUT
LOW-NOISE AMP
Figure 4-12. SSB Phase Noise Test Setup (Optional)
EQUIPMENT:
Mixer
HP 10514A
1 MHz Low Pass Filter
(See Figure 1-1)
Low-Noise Amplifier
HP 08640-60506
W/Power Supply HP 6216A
Spectrum Analyzer
HP 8566A
Oscilloscope
HP 1980B
LO
HP 8642B
PROCEDURE:
1.
Connect the instruments as shown in Figure 4-12. On the LO, select
external DC FM and 1 kHz FM deviation. Insure that the spectrum
analyzer center frequency calibration is accurate.
NOTE
The following symbols are used in the procedure:
fgp =
the RF output frequency where SSB phase
noise will be checked.
'OFF
the frequency offset from fRF where SSB
phase noise will be checked.
2.
Set the LO's frequency to f^p and the output level to +16 dBm. Set
the HP 8642 under test to f^p + fQFF
anc
^
t
^
ie o u t
P
u t
l
e v e
l
t o
"^0
dBm.
4-43