Table 3: Lowest temperatures at max. speed
rotor-number
220.72 V06
220.78 V05
221.54 V02
221.55 V02
221.12 V03
221.16 V03
221.17 V07
221.19 V02
221.18 V02
221.20 V02
221.22 V02
221.28 V02
220.87 V09
220.87 V10
221.23 V02
220.88 V09
220.92 V06
221.38 V01
All temperature indications refer to a room temperature of 23°C. By exceeding this value or direct solar radiation
to the centrifuge, these values can´t be kept up.
© Hermle Labortechnik GmbH
max. speed
n-max
5000 rpm
-3 °C
13500 rpm
17 °C
6000 rpm
-11,5 °C
6000 rpm
-8 °C
4500 rpm
-7 °C
4500 rpm
-6 °C
14000 rpm
9 °C
4500 rpm
-9 °C
9000 rpm
-5 °C
12000 rpm
5 °C
12000 rpm
-5 °C
12000 rpm
2 °C
15000 rpm
6 °C
15000 rpm
6 °C
18000 rpm
-2 °C
13500 rpm
1 °C
13500 rpm
2 °C
15000 rpm
-1 °C
Z326K_V3.15