Hermle Z 326 K Operating Manual

Hermle Z 326 K Operating Manual

Table top centrifuge
Table of Contents

Advertisement

Operating manual for
Table Top Centrifuge
Z 326 K
© Hermle Labortechnik GmbH
Z326K_V3.15

Advertisement

Table of Contents
loading

Summary of Contents for Hermle Z 326 K

  • Page 1 Operating manual for Table Top Centrifuge Z 326 K © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 3: Table Of Contents

    Preselection of speed / RCF-value ......................19 2.3.2 Preselection of running time ........................20 2.3.3 Preselection of brake intensity and acceleration ..................21 2.3.4 Pre-selection of temperature ........................22 2.3.5 Pre-cooling ............................... 22 Radius correction ............................23 Program ............................... 24 © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 4 Table 7 (part 1): Radius correction ........................XI Table 8 (part 2): Radius correction ........................XII Table 9 (part 3): Radius correction ........................XIII Table 10: Abbreviations used ......................... XIV Redemption form / Decontamination certificate ....................XV Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 5: Product Description

    From the resulting damage the company Hermle Labortechnik is not liable. 1.3 Brief description The unit type Z 326 K is a non refrigerated universal centrifuge, which we offer in two voltage variations 230V or 120V. The centrifuge can be used with swing-out and angle rotors.
  • Page 6: Operating And Display Elements

    / deceleration intensity time centrifugation time lid release quick short running start start centrifugation stop stop centrifugation prog calling stored programs store program store temp temperature indication Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 7: Lcd-Display

    „service“ „rcf“ h m s „accel“ „temperature“ Signs and indications of the centrifuge 1.6.1 General Instructions for disposal (see P. 34) Direction of rotation – clockwise rotation for the rotor drive Reference for loading rotors © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 8: Product-Nameplate (Example)

    PRODUCT DESCRIPTION 1.6.2 Product-nameplate (Example) Company address: Hermle Labortechnik GmbH, Siemensstr. 25, D-78564 Wehingen TYPE: Type designation of the product REF: Order no. of the product Serial number of the product Date of manufacture Manufacturer MAX. Drehzahl: max. allowed speed of the unit KIN.
  • Page 9: Warning And Information Signs

    EN 61010-2-020. • The HERMLE Z 326 K is explosion-proof and must therefore not be operated in explosion- endangered areas or locations. Centrifugation of flammable, explosive, radioactive, or such substances, which chemically react with high energy, is strictly prohibited. The final decision on the risks associated with the use of such substances is the responsibility of the user of the centrifuge.
  • Page 10: Following Rules Must Strictly Be Adhered To

    • Do not use accessories such as rotors and buckets, which are not exclusively approved by HERMLE Labortechnik GmbH, except commercially available centrifuge tubes made of glass or plastic. • Do not spin extremely corrosive substances, as they may cause material damages and impair mechanical resistance.
  • Page 11: Installation Of The Centrifuge

    PRODUCT DESCRIPTION 1.7 Installation of the centrifuge 1.7.1 Unpacking the centrifuge Model Z 326 K is supplied in a carton. Remove the strap retainer, open the carton and remove the centrifuge. The instruction manual must always be kept with the centrifuge! 1.7.2 Space requirements...
  • Page 12: Basic Adjustments

    All changed settings must be confirmed by the key „start“ (9). As an optical confirmation appears the word „store“ in the display „rpm/rcf“(A-1) - Only then the pre-selections are valid! (see figure 3) Figure 3 Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 13: Temperature Indication

    If you press the key „rpm/rcf“ (4) now, the word „On“ flashes and you can switch off the sound with the adjusting knob (1) (see figure 5). After you have stored the settings you changed to the normal program mode again by switch off the centrifuge for a short while. Firgure 5 © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 14: Volume Pre-Selection Of Sound Signal

    SonGo/Sound“. After pressing the key „rpm/rcf“ (4), you can select a song with the adjusting knob (1). (see figure 7). After you have stored the settings (see 1.8.1) you changed to the normal program mode again by switch off the centrifuge for a short while. Figure 7 Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 15: Keyboard Sound Turn On / Off

    „ON/BEEP“. After pressing the key „rpm/rcf“ (4), you can turn the keyboard sound (On) or (Off) with the adjusting knob (1). (see figure 8). After you have stored the settings (see 1.8.1) you changed to the normal program mode again by switch off the centrifuge for a short while. Figure 8 © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 16: Call Up Operating Data (By Skilled Or Service Engineer Only!)

    (1). The respective error codes appear in the display „rpm/rcf“ (A-1). Please look up in „Table 6: error messages“ (see APPENDIX S. X). Here as well you must shortly switch off the centrifuge for changing to the normal program mode again. Figure 9 Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 17: Operation

    Hold the rotor with one hand and secure the rotor to the shaft by turning the fixing nut clockwise. Tighten fixing nut with enclosed allen key (see figure 11) Figure 12 ATTENTION: For safety always ensure that rotor fixing screw is tightened before each run!! (see figure 11) © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 18: Loading Angle Rotors

    In principle swing out rotors may not be taken in operation until all buckets or racks are put into the rotor. The bolts at the rotor must be greased with the HERMLE vaseline (Order No. 41-5215). The sample tubes have to be filled evenly by eye and put into the drillings or tube racks. The weight difference of the loaded buckets should not exceed approx.
  • Page 19: Loading And Overloading Of Rotors

    If In case of any questions, please contact the manufacturer! 2.1.5 Removing the rotor Untighten the rotor fixing nut complete (2. screw over the stiff point) and lift the rotor vertical out of the centrifuge. (see figure 10 and 11) © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 20: Lid

    All with number marked passages refer to figure 17 ATTENTION: Don't grip your fingers between lid and device or locking mechanism when closing the lid! Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 21: Preselection

    20 xg and the maximum permissible centrifugal force of the rotor. See „table 4: max. speed and RCF-values for permissible rotor“ (see APPENDIX P. VIII). There are listed all important values. ATTENTION: Please also check the maximum permissible revolutions of your test tubes! (Producer Indication) © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 22: Preselection Of Running Time

    „time” (A-3). That continuous run can only be interrupted by pressing the key „stop” (10). The time countdown as soon as the set speed is reached. The display shows always the remaining running time. (see figure 19) All with number marked passages refer to figure 19 Figure 19 Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 23: Preselection Of Brake Intensity And Acceleration

    All with number marked passages refer to figure 20 See „table 5: acceleration and deceleration times“ (APPENDIX P. IX). There are shown the acceleration and deceleration times for the acceleration and deceleration stages 0 to 9 for permissible rotors. Figure 20 © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 24: Pre-Selection Of Temperature

    After the pre-set temperature is reached you can leave the pre-cooling run with the “stop” key (10). Depending on the inserted rotor the pre-cooling goes between approx. 10 and 20 min. All with number marked passages refer to figure 22 Figure 22 Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 25: Radius Correction

    As soon as you have set a radius correction the word „radius“ (M9) appears. This hint is as long visible as you put the radius correction back to 0 again. All with number marked passages refer to figure 23 Figure 23 © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 26: Program

    (M10) – the (xx) stands for the chosen program place. If all program numbers are occupied you can take an old number that is not necessary anymore and just put in the new parameters. Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 27: Recall Of Stored Programs

    To leave the program mode just press the key „prog“ (11). Then inside the display „time“ appears the word „programm". Set the display to „programm--“ (M10) with the adjusting knob (1). All with number marked passages refer to figure 27. © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 28: Starting And Stopping The Centrifuge

    2.6.2 The "STOP" key By the „stop“ key (10) (see figure 29) you can interrupt the run at any time. After pressing the key the centrifuge decelerates with the respective pre-selected intensity down to stand still. Figure 29 Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 29: Imbalance Detection

    Load the rotor as described in chapter 2.1.2 and 2.1.3. When inside the display “time” (A-3) the word “error” together with the number “02“ (see figure 30) appear, there could be following reasons: The imbalance switch is defective. Figure 30 © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 30: Maintenance

    Clean as well the accessories especially the sealings of the aerosol-tight rotors and insert bolts of swing out rotors. Afterwards lubricate the bolts or sealings with the recommended HERMLE vaseline - order no. 41-5215. Lubricants containing molycote and graphite are not allowed.
  • Page 31: Cleaning And Disinfection Of The Unit

    The aerosol tightness of rotors, rotor lids, buckets and caps has been tested and certified by the „TÜV Nord CERT GmbH, Certification Body Consumer Products, Essen (Germany)“ in accordance with Annex AA of IEC 61010-2-020. The certificates can be downloaded on our webpage www.hermle-labortechnik.de. Aerosol-tight rotors and buckets are marked with the label „aerosol-tight“.
  • Page 32: Disinfection Of Pp-Rotors

    Transparent rotor lids and caps made of PC or PP as well as rotors, tube racks and adapters of PP have a maximum operating time up to 3 years from first use. Condition for the operating time: Proper use damage-free condition, recommended care. Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 33: Trouble Shooting

    • Now turn the box spanner to the right side (clockwise) up to the limit. ATTENTION: Just turn to the limit, don´t tighten the nut. • Now open the lid of the centrifuge. • Switch the centrifuge on again, for go on working. Figure 31 © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 34: Description Of The Error Message System

    4.2.2 Description of the error message system The error message „error“ (M11) is shown in the „time“ (A-3) display (see figure 32). Detailed information about possible error messages are in „table 6: error messages" (see Appendix P.X). Figure 32 Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 35: Receipt Of Centrifuges To Repair

    Decontamination certificate at goods return delivery (see APPENDIX P. XVI) We reserve the right to accept contaminated centrifuges. Further on all costs occurred for the cleaning and disinfection of the units will go to the debit of the customer’s account. © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 36: Transport, Storage And Disposal

    RoHS II Compliance HERMLE Labortechnik GmbH, Siemensstraße 25, 78564 Wehingen, hereby declares and certifies that all components manufactured are RoHS II compliant according to the definition and restrictions given by the European Parliament Directive 2011/65/EC on the restriction of use of certain hazardous substances in electrical and electronic equipment.
  • Page 37: Appendix

    Table 7 (part 1): Radius correction ................... XI Table 8 (part 2): Radius correction .................. XII Table 9 (part 3): Radius correction ................. XIII Table 10: Abbreviations used ..................XIV Redemption form / Decontamination certificate ............. XV © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 38: Eg - Conformity Declaration

    Z 206 A; Z 233 M-2; Z 216 M; Z 306; Z 326; Z 366; Z 446; Z 513; Z 216 MK; Z 32 HK; Z 326 K; Z 366 K; Z 36 HK, Z 446 K, Z 513 K Einschlägige EG-Richtlinien / Normen...
  • Page 39 APPENDIX Table 1: Technical Data Manufacturer HERMLE Labortechnik GmbH, 78564 Wehingen Z 326 K Type Dimensions Width 40 cm Depth 70 cm Height 36 cm Weight without rotor 60 kg max. speed 18000 min max. volume 4 x 100 ml max.
  • Page 40 220.87 V09 15000 rpm 81,6 g 220.87 V10 15000 rpm 81,6 g 221.23 V02 18000 rpm 40,8 g 220.88 V09 13500 rpm 149,6 g 220.92 V06 13500 rpm 70,4 g 221.38 V01 15000 rpm 14 g Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 41 2 °C 221.38 V01 15000 rpm -1 °C All temperature indications refer to a room temperature of 23°C. By exceeding this value or direct solar radiation to the centrifuge, these values can´t be kept up. © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 42 221.28 V02 12000 rpm 15938xg 220.87 V09 15000 rpm 21381xg 220.87 V10 15000 rpm 21381xg 221.23 V02 18000 rpm 23545xg 220.88 V09 13500 rpm 17115xg 220.92 V06 13500 rpm 16708xg 221.38 V01 15000 rpm 15344xg VIII Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 43 1050 221.20 V02 221.22 V02 221.28 V02 220.87 V09 220.87 V10 221.23 V02 220.88 V09 220.92 V06 221.38 V01 in seconds Acceleration time Deceleration time from 0 min -> U from U -> 0 min © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 44 Timeout between control unit and frequency converter Other error frequency converter Overspeed Timeout between controler and RS232 interface Rotor is not allowed in this centrifuge Inserted rotor does not exist in the programm FALSE Rotor is not detected rotor no Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 45 705.015 14,0 -0,6 705.016 14,0 -0,6 13,9 -0,7 Angle rotor 220.87 704.004 -0,4 704.005 -1,1 Festwinkelrotor 221.54 V02 701.011 10,6 701.012 701.015 Festwinkelrotor 221.55 V02 708.019 10,7 708.003 10,3 708.004 10,6 701.011 10,2 701.012 701.015 © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 46 626.014 -4,9 626.015 11,6 -3,2 Swing out rotor 10,2 221.16 706.000 10,2 Angle rotor 221.17 704.004 -0,4 704.005 -1,1 Angle rotor 221.20 707.001 -0,7 707.002 -0,8 707.003 -0,3 707.004 -0,6 707.014 -0,9 707.000 -1,7 -0,1 Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 47 10,6 -0,6 707.002 10,4 -0,8 707.003 10,9 -0,3 707.004 10,6 -0,6 707.014 10,4 -0,8 707.015 10,4 -0,8 12,5 Angle rotor 221.19 10,9 701.000 -4,5 701.010 11,3 -1,2 701.011 12,2 -0,3 10,6 701.012 10,5 -2,0 XIII © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 48 APPENDIX Table 10: Abbreviations used Symbol/Abbreviations Unit Description U (=rpm) [min revolutions per minute RZB(=rcf) [x g] relative centrifugal force Polypropylen Polycarbonat accel acceleration decel deceleration prog program Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 49: Redemption Form / Decontamination Certificate

    These substances have reached the equipment/assembly? …………………………..  Yes  No Which one, if yes: ________________________________________________________________ Description of the measures for the decontamination of the listed parts: I confirm the proper decontamination: Company/dept ._____________ place and date: ____________________________________ Signature of the authorized person:________________________________________ © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 50 Z326K_V3.15 © Hermle Labortechnik GmbH...
  • Page 51 © Hermle Labortechnik GmbH Z326K_V3.15...
  • Page 52 HERMLE Labortechnik GmbH Siemensstraße 25 78564 Wehingen Tel: 0 74 26-96 22-17 Fax: 0 74 26-96 22-49 Email: vertrieb@hermle-labortechnik.de Internet: http://www.hermle-labortechnik.de Technische Änderungen vorbehalten. ©HERMLE Labortechnik GmbH Z326K_V3.15 © Hermle Labortechnik GmbH...

Table of Contents