When the "Information Display" is displaying "CP", the device
has been exposed to temperatures below -15° Celsius (5°F)
and will not operate properly . Set the device aside and allow it
to reach room temperature before using .
Note: Consecutive uses and high ambient temperatures
can cause the Interchangeable Head surface to become
warm to the touch. The device monitors the Head surface
temperature during each care cycle to ensure that the
Head maintains an optimum temperature. If the device
exceeds the optimum temperature limit, it will enter Heat
Protection Mode and the device will limit operation until
sufficient surface cooling has been achieved.
When the Information Display is displaying "HP", the device has reached its
optimum temperature limit and will not continue to operate until a cool down
period has been completed .
If your device enters Heat Protection Mode:
The Information Display will display "HP" and the device will beep 10 times and
immediately terminate the cycle .
Step 1: Set the device back on the Charging Stand and allow cooling .
Step 2: Once "HP" is no longer visible in the Information Display, the device
is ready for use .
Step 3: Resume normal operation as outlined on page 9 of this manual .