ENGLISH (M5)-5 9/18/02 3:20 PM Page 10
Error Symbol
If you continue to experience difficulty obtaining a reading, contact your nearest
OMRON representative or dealer.

Error Indicators

Cause
Cuff under-inflated.
Movement during
measurement
No pulse or blood
pressure reading due to
very weak artery pulse.
Cuff over-inflated
(more than 300 mmHg).
Batteries have run down.
10
Correction
Wait 2-3 minutes and
take another measurement.
Repeat steps listed under
"How to Take a Reading"
but manually inflate the cuff
40 mmHg higher than the
previous inflation value.
Carefully read and repeat
steps listed under "How to
Take a Reading."
Replace all four "AA"
batteries.