Calibration Adjustments:
NOTE: Do not deviate from the calibration adjustment sequence
that follows. This will ensure that adjustments to be stored in
EEPROM go to the correct locations.
1. Perform meter-mode adjustments and verifications per Table 3 (use Table
5 for RTD-Thermistor models).
2. Hit OPR key (Flashing 'RTD' annunciator turns off).
3. Install J1 jumper.
4. Set unit to CALIB mode, RTD-100, °F.
5. Switch current source to OPERATE mode.
6 Perform calibrator-mode adjustments and verifications per Table 4 (use
Table 6 for RTD-Thermistor models).
7. Turn off unit. Remove tape holding battery.
8. Re-install back cover. Resistance-function calibration is complete.
Table 3: Calibration of Meter-Mode Resistance Functions
(RTD-100/1000)
Sensor
Step
Select
1
RTD-100
2
RTD-100
3
RTD-100
4
RTD-1000
5
RTD-1000
6
1,000Ω
7
1,000Ω
8
10,000Ω
9
10,000Ω
* exclusive of noise.
** adjust as necessary to obtain steps 2, 5.
30
Unit
Input
Adjust
390.26Ω
P1**
100.00Ω
18.49Ω
3,902.6Ω
P1**
1000.0Ω
900.00Ω
P1
1.00Ω
9,000.0Ω
P1
10.0Ω
Unit
Verify
Reading*
1562.0°F±0.5°F
32.0±0.1°F
–328.0±0.2°F
1562.0°F±0.5°F
32.0+0.1°F
1.00±0.10Ω
10.0±1.0Ω
900.00Ω
9000.0Ω