Setting Input Range And Recording Span (Range) - Omega RD260A User Manual

Omega products, inc recording equipment user manual
Table of Contents

Advertisement

6.1 Setting Input Range and Recording Span (RANGE)

NOTE
6 - 2
One of the following parameters can be set for every channel:
SKIP
Prevents the specified channel from being measured, recorded and displayed .
VOLT
Measures and records DC-voltage.
TC
Measures and records temperatures using Thermocouple.
RTD
Measures and records temperatures using Resistance Temperature Detector .
DI
Accepts Digital Input (contact/voltage level) and performs ON/OFF
recording.
DELT
Performs difference computation between two channels which have same
input range.
SCL
Performs scaling on voltage measurements.
Scaling results in appending a different scale to the measured values.
SQRT
Takes the square root (
Restrictions apply to the measurable types of input depending on the type of
recorder used. Check your recorder type as indicated on the nameplate (see
page 1-4 of the instruction manual for details on the nameplate), and then refer
to the following table.
Model Code
Measurable Type of Input
SR100x-B-1
Voltage (VOLT), thermocouple (TC), contact (DI), SCL, SQRT
SR100x-B-2
Resistance temperature detector (RTD)
) of DC-voltage measured data.

Advertisement

Table of Contents
loading

Table of Contents