Page 1
Gas Detector Biosystems PHD6 Reference Manual Sperian Instrumentation 651 South Main Street Middletown, CT 06457 800 711-6776 860 344-1079 Fax 860 344 – 1068 12MAR2009 Part Number 13-322 Version 2.01 http://www.biosystems.com...
Page 2
BIOSYSTEMS PHD6 PERSONAL PORTABLE GAS DETECTORS HAVE BEEN DESIGNED FOR THE DETECTION AND MEASUREMENT OF POTENTIALLY HAZARDOUS ATMOSPHERIC CONDITIONS IN ORDER TO ASSURE THAT THE USER IS PROPERLY WARNED OF POTENTIALLY DANGEROUS ATMOSPHERIC CONDITIONS, IT IS ESSENTIAL THAT THE INSTRUCTIONS IN THIS REFERENCE MANUAL BE READ, FULLY UNDERSTOOD, AND FOLLOWED.
1.4.7 Latching alarms 1.4.8 Fault detection Other electronic safeguards Sensors Optional sample draw pump 1.7.1 Special precautions when using the PHD6 pump Data storage 1.8.1 Black box data recorder 1.8.2 Event logger PHD6 design components 1.10 PHD6 standard accessories 1.10.1 Alkaline PHD6 detectors 1.10.2...
Page 4
Cleaning and replacing PID components Sample probe assembly 6.5.1 Changing sample probe filters 6.5.2 Changing sample probe tubes (wands) PHD6 Pump Maintenance 6.6.1 Replacing pump filters PPENDICES Appendix A Toxic gas measurement – Warning, Danger, STEL and TWA alarms Warning and Danger Alarms...
Operating Temperature and Humidity Limits The Biosystems PHD6’s operating temperature range is printed on the label on the back of the instrument. Use of Sperian Gas Detectors outside of the instrument’s specified operating temperature range may result in inaccurate and potentially dangerous readings.
Page 6
Sperian part number 54-49-103-5. The PHD6 must be located in a non-hazardous location during the charging cycle. Charging the PHD6 in a hazardous location may impair intrinsic safety. PHD6 rechargeable battery packs are supplied with Panasonic CGR18650D Lithium-Ion batteries. The Li-Ion batteries in the battery packs may not be replaced by the user.
PHD6 are contained in the Calibration procedures are discussed in detail subsequent chapters of this manual. in Chapter 4. The PHD6 can utilize a variety of sensor types to detect atmospheric contaminants including electrochemical sensors, PID (Photo Ionization Detector) sensors, NDIR (Non-Dispersive Infra- Red Absorbance) sensors and catalytic hot-bead LEL sensors.
1.4.3 Sensor over range alarms specific sensor. When an alarm set point is The PHD6 will go into alarm if a sensor is exceeded a loud audible alarm sounds, and the exposed to a concentration of gas that exceeds bright red LED alarm lights flash.
The PHD6 can be configured to simultaneously and solvents. If the PHD6 is being used in an monitor oxygen, combustible gases and vapors, atmosphere that may contain concentrations of...
"MODE" button. The MODE button is To extract the information from the black box data used to turn the PHD6 on and off as well as recorder, the PHD6 must be returned to Sperian. to control most other operations, including...
Once 1.11 PHD6 kits the memory is filled, the PHD6 will start to write new data over the oldest data in order to PHD6 detectors may also be purchased as part conserve that most recent data.
Once the PHD6 has completed the start up hazard alarms, see section 2.4. sequence, the current gas readings screen will be After the alarm screens, the PHD6 will show that shown. The status bar at the bottom of the “Starting Session, Resetting Averages” followed display shows time plus calibration, pump and by the calibrations status screen.
“None” and the triangular warning symbol at with the remaining sensors. the Loading Sensors screen. The PHD6 also tests the lamp in the PID sensor at regular intervals during normal operation. If the PHD6 determines that the lamp has gone out,...
Removal of the alkaline 1. If the PHD6 is turned off, hold the MODE button down for about 5 seconds until batteries from the battery pack in a hazardous area may impair intrinsic safety. “Communication Mode” is shown. If the PHD6 is on already, proceed to step 2.
Special Instructions for NDIR The PHD6 can also be used to sample remote sensors locations with either the hand-aspirated sample- Two NDIR sensors are available for the PHD6: draw kit, or with the motorized sample draw One for the detection of carbon dioxide (CO pump.
6. Note the gas measurement readings. tabs on the pump into the corresponding slots on CAUTION: Hand-aspirated remote sampling the back of the PHD6. Once the pump is in only provides continuous gas readings for the position over the sensors, tighten the knurled...
PHD6 Calibration When the pump is active and The accuracy of the PHD6 should be verified on functioning properly, the moving a regular basis. Verification can be as simple as pump icon is shown on the lower performing a bump test, which is described below status bar on the display.
Appendix B. to section 4.3 and perform the gas Functional (Bump) testing calibration. The accuracy of the PHD6 may be verified at any The Canadian Standards Association Note: time by a simple functional (bump) test. (CSA) requires combustible gas sensors to...
0 without pressing the the instrument will display MODE button. the following screen. Note If the PHD6 still fails to calibrate after this that the sensor(s) that fail procedure is attempted, contact Sperian. the zero calibration are shown (in this case, CO) 4.2.3...
“no GAS”. See section 5.2.4 under Gas Values for more information on the O2 sensor check. When gas calibration is due, the PHD6’s display will show the warning symbol while intermittently A sensor that cannot be displaying the calibration bottle in the gas calibrated or is found to be out of tolerance readings screen.
, leave the calibration gas on and press the MODE button three times within two seconds to To access the Basic Menu, with the PHD6 on and initiate the zero calibration sequence. Press the current gas readings screen shown, hold the...
PHD6’s functionality. NOTE: Changes made in the Main Menu can have a direct affect on the PHD6’s functionality and should only be made by those who are trained in proper gas detection and monitoring techniques.
Configuration Menu the oxygen sensor check • Security Beep (enable or disable) may result in the failure to Once enabled the PHD6 will emit a short recognize an oxygen- audible beep and give a short flash on the deficient atmosphere.
Information Menu the upper housing by pulling up and away. • Versions (view instrument serial number, CAUTION Always turn the PHD6 off prior to software version, and time and date of removing the battery pack. Removal of the instrument manufacture)
30 degrees Celsius (86 degrees Fahrenheit). Li-Ion batteries may suffer 1. If the PHD6 starts and the battery icon if full, deterioration resulting in damage to the internal then the battery is fully charged and may be components when stored at high temperatures.
3. Baseline becomes generally unstable. swab starts to squeak. 4. Baseline shifts when the instrument is in 6. Reassemble the sensor and the PHD6. See motion. steps 3-8 above in the directions to replace When to clean the PID lamp the stack or lamp.
PHD6 Pump Maintenance contamination. PHD6 pumps are fairly maintenance free with the CAUTION: Never perform remote sampling exception of the replacement of the pump filters without the sample probe and hose on a regular basis.
Appendices alarm levels in the PHD6 are less than or equal to the OSHA-assigned ceiling levels for both CO and H Never enter an environment even momentarily Appendix A Toxic gas measurement – when concentrations of toxic substances exceed Warning, Danger, STEL and TWA the level of either the Warning or the Danger Alarm.
One of the most common questions “Instrument Products” group of this trigger a resumption of daily that we are asked at Sperian organization has been very active checking. Instrumentation is: “How often in developing a protocol to clarify...
Page 31
Appendix D Electrochemical Toxic Sensor Cross-Sensitivity The table below provides the cross-sensitivity response of the PHD6 electrochemical toxic gas sensors to common interference gases. The values are expressed as a percentage of the primary sensitivity, or the reading of the sensor when exposed to 100ppm of the interfering gas at 20ºC.
Page 32
Instrument’s Reference Manual are not covered by the Sperian warranty. The obligation of Sperian under this warranty is limited to the repair or replacement of components deemed by the Sperian Instrument Service Department to have been defective under the scope of this standard warranty.
Need help?
Do you have a question about the PHD6 and is the answer not in the manual?
Questions and answers