Last Zero - GSE 370 Technical Reference Manual

Precision counting scale
Hide thumbs Also See for 370:
Table of Contents

Advertisement

DISPLAY READS
2.
Key in 54321 [SAMPLE/Enter].
DISPLAY READS
3.
Remove any load on the scale.
DISPLAY READS
4.
Press [SAMPLE/Enter] to establish zero.
DISPLAY READS
5.
Pause for motion delay.
DISPLAY READS
6.
Place a 100lb test weight on scale.
DISPLAY READS
7.
Enter 100.
DISPLAY READS
8.
Press [SAMPLE/Enter] to establish span.
DISPLAY READS
9.
Pause for motion delay.
DISPLAY READS
10. Press [SAMPLE/Enter] to accept calibration.
DISPLAY READS
11. Press [SAMPLE/Enter] to save calibration.
DISPLAY READS
12. Press [SAMPLE/Enter] to exit calibration.
DISPLAY READS
13. Remove the calibration weigh
DISPLAY READS
L
Z
AST
ERO
The Last Zero procedure allows recalibration of the weighing device using an existing test load.
This is especially beneficial when checking high capacity applications such as tank weighing to
minimize the task of placing and removing test weights.
Establish gross zero before entering setup or calibration!
Last Zero Calibration With Weight Already Applied Example (370):
1. Remove any load on the scale.
DISPLAY READS
2. Press [ZERO] to zero the scale.
DISPLAY READS
3. Apply a 10000 lb test weight to verify calibration.
DISPLAY READS
4. Press [ZERO] + [SELECT].
DISPLAY READS
5. Press [ZERO] [PRINT] [UNITS] [TARE].
DISPLAY READS
6. Press [SELECT].
DISPLAY READS
Setup
Fast ~ Cal~First ~ Zero? ~ -0.26
First ~ Zero? ~ -0.42
0.00
Enter ~ Load ~ 0.00
Enter ~ Load ~ 99.66
100
100.00
Cal ~ Good? ~ 100.00
Enter ~ =Stor
Enter ~ =End
100.00
100.00
10
00
9970
Setup
Fast ~ Cal~First ~ Zero? ~ 9930
Last ~ Zero? ~9930
63