11. OFFSET ADJUSTMENT

11-1 Type K/J/T/E/R/S offset adjustment
1)Set the function to Type K ( or other type J/E/R/T/S ).
2)Insert the probe to the T1 input socket ( 3-9, Fig. 1 )
3)Pressing " Offset button " ( 3-8, Fig. 1 )
continuously at least two seconds then release,
the display will show :
measuring
value
* Use the "
Fig. 1 ) to adjust the desiring value on right bottom
display.
* Pressing " Enter button " ( 3-4, Fig. 1 ) once, the
adjustment value will save into memory then return
to normal measuring screen and finish the offset
adjustment procedures.
* The above offset adjustment for Type K/J/T/E/R/S is
valid for T1, T2, T3, T4 at the same time.
7-2 Pt 100 ohm offset adjustment
1)Set the function to Pt 100 ohm.
2)Insert the Pt 100 ohm probe to the PT1 ( PT2 ) input socket.
9600
No parity
8 Data bits
1 Stop bit
SEt
oFS
23.7
23.7
adjust
value
button "( 3-5, Fig. 1 ) "
* If not insert probe
22