Table 12-2
Ports and cabling – Docked (continued)
Port name on
EUT
LAN
VGA

Electromagnetic emissions

The HP ElitePad 1000 G2 Healthcare Tablet is intended for use in the electromagnetic environment specified below. The customer or
the user of the HP ElitePad 1000 G2 Healthcare Tablet should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
WARNING!
could result in improper operation. If such use is necessary, this equipment and the other equipment should
be observed to verify that they are operating normally.
The HP ElitePad 1000 G2 Healthcare Tablet is intended for use in the electromagnetic environment specified below. The customer or
the user of the HP ElitePad 1000 G2 Healthcare Tablet should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Cable description
Qty
(or reason for no
cables)
Ethernet
1
VGA
1
Guidance and manufacturer's declaration—electromagnetic emissions
Use of this equipment adjacent to or stacked with other equipment should be avoided because it
Guidance and manufacturer's declaration—electromagnetic immunity
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Length as
Max length
tested (m)
(m)
2.0
>3.0
1.0
>3.0
Compliance
Group 1
Class B
Class D
Class Complies
Compliance level
± 8 kV contact
± 2, 4, 8, 15 air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 0.5 kV, ± 1 kV line(s) to
line(s)
Shielded?
Termination box ID &
port name
(Y/N)
No
Switch
No
Monitor
Electromagnetic environment—guidance
The ElitePad 1000 G2 uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
The ElitePad 1000 G2 is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Electromagnetic environment—
guidance
Floors should be wood, concreate, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
EN 60601-1-2: 2007