If you do not resolve the problem, contact your
WHAT YOU SEE
WHAT IT MEANS
Test result is above
33.3 mmol/L.
CAUTION: Glucose levels above 13.9 mmol/L may
indicate a potentially serious medical condition.
Test result is
below 0.6 mmol/L.
CAUTION: Glucose levels below 2.8 mmol/L may
indicate a potentially serious medical condition.
The display goes
•
Meter will turn off
after 3 minutes
blank after the
of inactivity. This
meter is turned on.
feature prolongs
your battery life.
•
The battery
may need to be
replaced.
Results are not dis-
This may affect
played in expected
the way you see
units of measure
your results.
(mmol/L or mg/dL).
26
Need Help? Call your local Bayer Diabetes Care Customer Service
81656517_CntrTS_UG_EN_RU_FpBp_v2.indd 26-27
81656517_CntrTS_UG_EN_RU_FpBp_v2.indd 26-27
WHAT YOU SHOULD DO
•
Wash your hands and the test
site. Repeat the test using a
new test strip.
•
If your result is still "HI,"
contact your physician or
healthcare professional
immediately.
•
Repeat the test using a new
test strip.
•
If your result is still "LO,"
contact your physician or
healthcare professional
immediately.
•
Try turning the meter on again
by pressing
. If display
goes blank before 3 minutes,
replace battery.
See page 28.
Contact Customer Service.
local Bayer Diabetes Care Customer Service.
WHAT YOU SEE
WHAT IT MEANS
Some display seg-
Possible meter
electronics failure.
ments do not appear
when the meter
initially turns on.
Control test result
•
Test strip is deterio-
rated due to expo-
is out-of-range (too
sure to humidity or
high or too low).
heat.
•
Control solution
is past expiration
date or is past the
open use date.
•
Test strip is past
expiration date or
is past the open
use date.
•
Incorrect control
solution was used.
•
Control solution
is not at room
temperature.
•
Control solution is
not mixed well.
www.diabetes.bayer.com
WHAT YOU SHOULD DO
Press and hold
while the
meter is off. Compare meter dis-
play with the picture on page 3.
If the display is still not correct,
call Customer Service.
•
Run another control test with
a new test strip.
•
Check all expiration dates.
Do not use expired testing
materials.
•
Make sure you are using
C
TS control solution.
ONTOUR
•
Wait 20 minutes until control
solution is at room tempera-
ture. Try again.
•
Gently rock the control bottle
to ensure the control solution
is mixed well.
•
If result is still out-of-range,
retest with a new test strip
and control solution.
•
If result is still out-of-range,
call Customer Service.
27
12/10/13 12:05 PM
12/10/13 12:05 PM