For correct use of the device refer to the following chart. The
chart provides important information about the recommended
operating environments, or operating conditions.
Specifications
Operating
32º F to 104º F (0º C to 40º C)
Temperature
Operating
10% - 95% at an 82.4° F (28° C) dew point
humidity
Transport/
-13°F to 158°F (-25°C to 70°C)
Storage
Humidity: Up to 95% Non-Condensing
Temperature
-4°F to 122°F (-20°C to 50°C) for battery packs
electrical
Use no extension cords. Use no electrical outlets con-
trolled by a switch.
Altitude
0 – 13,123 feet (0 - 4,000 meters)
Placement
DO NOT block the air inlet (top back cover) or the exhaust
vent (back bottom right). Place device a minimum of 3
inches (7.5 cm) away from walls, draperies, furniture, etc.
environment Must be smoke, pollutant, and fume free.
Operating
24-hours a day when connected to an external AC or DC
Time
power source.
Note: If the Oxygen Concentrator has been stored for an extended period
of time outside its normal operating temperature range, the unit should
be allowed to return to normal operating temperature before being
turned on. (Refer to the Specifications section in this manual.)
Proper Placement of the eQuinox
Pick a location for the device that avoids the intake of smoke,
fumes and pollutants. Correct positioning of the device should
allow intake of air through the air inlet filter at the top rear of
the cabinet. It should also allow exhaust air to freely leave the
exhaust vent at the bottom right of the device.
Place the device so that the alarms can be heard.
Position the oxygen supply tubing in such a way that it does not
kink or occlude.
WARNING: DO NOT USe YOUR OXYGeN CONCeNTRATOR
IN The PReSeNCe OF FLAMMABLe GASeS. ThIS CAN
ReSULT IN RAPID BURNING CAUSING PROPeRTY DAMAGe,
BODILY INJURIeS OR DeATh. USe NO OIL, GReASe, OR PeTRO-
LeUM-BASeD OR OTheR FLAMMABLe PRODUCTS WITh The OXY-
GeN-CARRYING ACCeSSORIeS OR The OXYGeN CONCeNTRATOR.
ONLY WATeR BASeD, OXYGeN COMPATIBLe LOTIONS OR SALVeS
ShOULD Be USeD. OXYGeN ACCeLeRATeS The COMBUSTION OF
FLAMMABLe SUBSTANCeS.
WARNING: ThIS DeVICe SUPPLIeS hIGh-CONCeNTRATION OXYGeN
ThAT PROMOTeS RAPID BURNING. DO NOT ALLOW SMOKING OR
OPeN FLAMeS WIThIN The SAMe ROOM OF (1) ThIS DeVICe, OR (2)
ANY OXYGeN-CARRYING ACCeSSORY. FAILURe TO OBSeRVe ThIS
WARNING CAN ReSULT IN SeVeRe FIRe, PROPeRTY DAMAGe, AND/
OR CAUSe PhYSICAL INJURY OR DeATh.
Note: When the automobile in which you are using the Oxygen Con-
centrator unit is turned off, disconnect and remove the device from the
automobile. Do not store the Oxygen Concentrator in a very hot or cold
automobile or in other similar, high-or low-temperature environments.
DO NOT leave the Oxygen Concentrator or the Power Supply plugged
into the vehicle if the ignition is in the off position. Doing so may drain
the vehicle's battery.
CAUTION: Do not operate unit in a restricted or confined
space (i.e., a small case or handbag) where ventilation
can be limited. This can cause the Oxygen Concentrator
to overheat and impair performance. Do not allow either the air
intake or the air outlet vents to become blocked. This can cause the
Oxygen Concentrator to overheat and impair performance.
eQuinox (Model 4000) User Manual — 9