Ch.4 Checking Image Quality
1
2
Be sure to touch the [Save] button after inputting values.
Failure to save the values will result in printing of a calibration pattern in which input
values are not reflected.
4
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
4
Input the measured density values.
1)Touch the [NEXT] button on the tray selection/print
screen.
- The density value input screen will be displayed.
2)Input the density values measured in step-3.
3)Touch the [SAVE] button after completion of input.
3
4)Touch the [NEXT] button.
- The calibration pattern screen will be displayed.
5)Touch the [PRINT] button.
- A calibration sheet will be printed.
5
< 146 >