Installing The Hardware; Device Self-Calibration - National Instruments Network Device DAQ S User Manual

National instruments network device user's manual
Table of Contents

Advertisement

Chapter 1
Getting Started

Installing the Hardware

Device Self-Calibration

You can also programmatically self-calibrate your device with NI-DAQmx,
Note
as described in Device Calibration in the NI-DAQmx Help or the LabVIEW Help in
version 8.0 or later.
NI 6124/6154 User Manual
The DAQ Getting Started Guide contains non-software-specific
information about how to install PCI and PXI Express devices, as well as
accessories and cables.
NI recommends that you self-calibrate your S Series device after
installation and whenever the ambient temperature changes.
Self-calibration should be performed after the device has warmed up for the
recommended time period. Refer to the device specifications to find your
device warm-up time. This function measures the onboard reference
voltage of the device and adjusts the self-calibration constants to account
for any errors caused by short-term fluctuations in the environment.
Disconnect all external signals when you self-calibrate a device.
You can initiate self-calibration using Measurement & Automation
Explorer (MAX), by completing the following steps.
1.
Launch MAX.
2.
Select My System»Devices and Interfaces»NI-DAQmx
Devices»your device.
3.
Initiate self-calibration using one of the following methods:
Click Self-Calibrate in the upper right corner of MAX.
Right-click the name of the device in the MAX configuration tree
and select Self-Calibrate from the drop-down menu.
1-2
ni.com