Performing A Control Solution Test - Abbott FreeStyle Freedom Lite Owner's Booklet

Hide thumbs Also See for FreeStyle Freedom Lite:
Table of Contents

Advertisement

Performing a Control Solution Test

1.
Check control solution expiry date.
IMPORTANT: Do not use control solution past expiry date. Discard
control solution 3 months after opening or on the expiry date printed on
the bottle whichever comes first. (Example: open April 15, discard July
15; write the discard date on the side of the bottle.)
2.
Remove test strip from vial.
3.
Insert test strip to turn meter on.
Note: The meter turns off after 2 minutes of inactivity.
Remove and reinsert the unused test strip to restart
the meter.
If the meter does not turn on go to Troubleshooting section.
23
ART22311-101_rev-B.indd 23-24
www.AbbottDiabetesCare.com
Top
Printed side up
Sample areas
4.
Confirm System Check Screen.
This display appears when you turn on the meter. Do not use the meter if
the display check screen does not exactly match the example. Contact
Customer Care.
See The FreeStyle Freedom Lite Meter and How It Works section for more
information.
If the system check screen does not appear, go to Troubleshooting section.
5.
Apply and hold control solution to 1 sample area of the test strip until:
You see short lines moving clockwise on the display or
You hear a beep
This indicates the test strip obtained enough control solution and the meter is
processing. If the test does not start after applying the control solution, go to
Troubleshooting section.
IMPORTANT: If a short line does not appear after 5 seconds, the
sample may be too small. Do not add control solution to both sides
of the strip. The test strips are for single use only. Do not reuse test
strips. Dispose of used test strips.