Abbott FreeStyle Freedom Lite Owner's Booklet page 12

Hide thumbs Also See for FreeStyle Freedom Lite:
Table of Contents

Advertisement

Message
What It Means
There may be a problem
with the test strip
Meter error
Very high blood glucose
higher than 27.8 mmol/L
HIGH control solution
was applied when the
temperature was too
cold. (Applies only to
control solution labeled
as HIGH).
17
ART22311-101_rev-B.indd 17-18
What To Do
1. Follow your healthcare professional's advice to
treat high blood glucose if you have symptoms
such as:
Fatigue
Thirst
Excess urination
Blurry vision
2. Perform a control solution test using a new
test strip. If the test results are within the range
printed on the test strip vial, retest using blood
and a new test strip.
3. If the control solution result is out of range or the
error reappears, contact Customer Care.
www.AbbottDiabetesCare.com
Message
What It Means
Incorrect test procedure.
For example, putting
blood on the test strip
before inserting it into
the meter or applying
blood before
and
appear
There may be a problem
with the test strip
Meter error
There may be a problem
with the test strip
Meter error
What To Do
1. Wait for
and
to appear before applying
blood or control solution.
2. Perform a control solution test using a new
test strip. If the test results are within the range
printed on the test strip vial, retest using blood
and a new test strip.
3. If the control solution result is out of range or the
error reappears, contact Customer Care.
1. Perform a control solution test using a new
test strip. If the test results are within the range
printed on the test strip vial, retest using blood
and a new test strip.
2. If the control solution result is out of range or the
error reappears, contact Customer Care.
18
8/28/14 11:30 AM

Hide quick links:

Advertisement

Table of Contents
loading